Elekta, Inc.: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.

Recall number
Z-1339-2026
Recalling firm
Elekta, Inc.
Product
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.
Status
Open, Classified
Initiated
2025-12-12
Posted
2026-02-10
Terminated
not on file
Reason
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
Root cause
Use error
Action
On 12/12/2025, the firm sent an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter via email to customers informing them that when Defining a new stereotactic reference in an examination for a frame treatment while failing to first change an already defined obsolete stereotactic reference (representing a non-current frame fitting) to a pre-plan reference, may result in a treatment plan that irradiates the wrong patient location if the resulting inconsistent alignment between the stereotactic references is not detected by the user. Customers are instructed to: If the stereotactic reference for a frame treatment must be changed prior to plan approval (e.g. due to a change of frame fitting), make sure to first change the already defined obsolete stereotactic reference to a pre-plan reference before defining the new stereotactic reference. By so doing, the treatment examination is automatically updated according to the new stereotactic reference when it is defined. Do not combine the use of a stereotactic CBCT reference with other stereotactic references for treatments with a frame. This specific recommendation for CBCT is motivated by the increased likelihood of failure for users to detect an inconsistent stereotactic CBCT reference relative to other stereotactic references, compared to MR and CT image studies, which are typically displayed and explored throughout the planning process. Post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. " Advise the appropriate personnel, working with this product, on the content of this letter. For Questions - Contact Elekta Care Support at 855-693-5358 or email Application.Support.NA@elekta.com
Quantity
not on file
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
Lot, serial or model codes
Lot Code: 07340048311236 (LGP 11.1)
510(k) numbers
["K151666"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register