Siemens Medical Solutions USA, Inc: Artis Zee (material # 10094135, 10094137, 10094141) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee and Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Recall number
Z-1339-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis Zee (material # 10094135, 10094137, 10094141) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee and Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Status
Terminated
Initiated
2019-04-09
Posted
not on file
Terminated
2020-04-23
Reason
In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Root cause
Software design
Action
On April 9, 2019, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that ARTIS systems equipped with certain motor control units may encounter with hardware failure, and planned procedures may need to be terminated and performed on an alternative x-ray system. The firm stated their service organization would contact customers shortly to arrange a date to perform a software update to correct the issue. For an earlier appointment or if you have any questions, you may contact the firm at 1-800-888-7436. Customers were asked to promptly notify and instruct all staff at the organization who need to be aware of the notice. Additionally, if you further distributed the product, please identify your customers and notify them at once of this recall.
Quantity
459 (116 US)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL,AR, AZ, CA, FL, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NV, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, VA , WA WI and WV. and Foreign countries.
Lot, serial or model codes
Serial Numbers 138260 138221 138259 138250 138268 138266 124286 125245 148402 148401 148384 148390 147249 148393 148411 148400 148358 148361 148397 148392 148382 148363 148375 155237 155203 155226 155227 155235 131235
510(k) numbers
["K073290"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register