Skip to the content

FDA Medical Device Recalls (openFDA)

Celsion Corporation: Prolieve Thermodilatation System, model number M0068808000, labeled in part***Distributed by Boston Scientific Corporation, Natick, MA Manufactured by Celsion Corporation, Columbia, MD 21046***

Recall number
Z-1339-04
Recalling firm
Celsion Corporation
Product
Prolieve Thermodilatation System, model number M0068808000, labeled in part***Distributed by Boston Scientific Corporation, Natick, MA Manufactured by Celsion Corporation, Columbia, MD 21046***
Status
Terminated
Initiated
2004-07-19
Posted
2004-08-11
Terminated
2005-03-31
Reason
Software controlling a medical device for patient treatment may malfunction and cause thermal injury to patient.
Root cause
Other
Action
Celsion Corporation ceased distribution of the device 6/25/04 and implemented a field correction by company representatives at end user facilities on 6/25/04 consisting of computer clock reset. The firm subsequently notified end users on 7/19/04 by certified letter, which included a warning sticker to be placed on the device control screen. The notification and sticker instructed users not to use the device if there is a possibility of the computer clock transitioning through midnight during the procedure.
Quantity
28 devices
Distribution
Devices were distributed to 19 medical facilities and physician end users nationwide by the recalling firm''s contract distributor.
Lot, serial or model codes
CEL0204AK CEL0204BK CEL0204CK CEL0204DK CEL0204EK CEL0204FK CEL0204GK CEL0204HK CEL0204IK CEL0204JK CEL0604AK CEL0604BK CEL0604CK CEL0604DK CEL0604EK CEL0604FK CEL0604GK CEL0604HK CEL0604IK CEL0604JK CEL0604KK CEL0604LK CEL0604MK CEL0604NK CEL0604OK CEL0604PK CEL1904AK CEL1904BK
510(k) numbers
not on file
PMA numbers
["P030006"]
Product code
MEQ
Firm FEI
1122126
Firm city
Columbia
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register