Skip to the content

FDA Medical Device Recalls (openFDA)

Rayner Intraocular Lenses Ltd: RayOne Preloaded Hydrophilic Acrylic IOL Injection Systems containing one MICS injection system with Intraocular Lens: (1) Outer carton labeled as Rayner RayOne EMV US, REF RAO200E, SE: +19.00D, Sph: +19.00D; and (2) Inner primary device packaging labeled as Rayner RayOne EMV US, REF RAO200E, SE: +21.00D, Sph: +21.00D.

Recall number
Z-1338-2022
Recalling firm
Rayner Intraocular Lenses Ltd
Product
RayOne Preloaded Hydrophilic Acrylic IOL Injection Systems containing one MICS injection system with Intraocular Lens: (1) Outer carton labeled as Rayner RayOne EMV US, REF RAO200E, SE: +19.00D, Sph: +19.00D; and (2) Inner primary device packaging labeled as Rayner RayOne EMV US, REF RAO200E, SE: +21.00D, Sph: +21.00D.
Status
Open, Classified
Initiated
2022-05-13
Posted
not on file
Terminated
not on file
Reason
The outer package is mislabeled and the package contains a different IOL strength.
Root cause
Under Investigation by firm
Action
The recalling firm sent two emails to the importer warehouse on 5/12/2022 instructing them to freeze all units of the affected lots from further distribution, provide the number of units in inventory, and to provide the accounts who have been sent the units so they can track them. On 5/13/2022, the recalling firm sent an email to their sales team requesting all unused units from the affected batch be returned to a designated warehouse (the importer, which was listed as a direct account). The email included an attached letter and a response form that the sales team was to send to any of their accounts who are affected. The email noted that Rayner would cover all return costs and replace the affected stock. The attached customer letter was dated 5/13/2022 requesting return of all unused units of the affected batch to the address listed in the letter. The letter noted that an official letter will follow subject to FDA approval, however, in the interim, the customer was requested to complete and returned the attached response form. The response form was to be completed with the lot number of the IOL, whether it was implanted, the date implanted, was there a refractive error, whether it is being returned, and the date of return. The facility is to indicate they have read and understood the contents of the letter and that all affected persons have been notified of the field action. A second letter was issued to the customers dated 5/24/2022 via email on the same date. The customer was requested to complete and return the attached response form at their earliest convenience if not already done. The email also asked for additional information be provided if the IOL has been implanted, such as (1) confirm the patient's post-operative outcome; (2) if applicable, confirm the degree of the refractive error; (3) does the refractive error require correction; and (4) if the refractive error requires correction, is surgical exchange expected. The attached letter expla
Quantity
162 IOLs
Distribution
US Nationwide distribution in the states of AZ, CA, FL, GA, IN, MI, MN, MO, ND, NJ, OK, TN, and TX.
Lot, serial or model codes
Outer carton +19.0 D - Batch #012183745, exp. 1/17/2024, Serial numbers 01-06, 10, 21, 24-45, 48-53, 58-60, 61-65, and 68-81, UDI (01)05029867006685. Outer carton +21.0 D - Batch #012183845, exp. 1/17/2024, all serial numbers, UDI (01)05029867006722.
510(k) numbers
not on file
PMA numbers
["P060011S025"]
Product code
HQL
Firm FEI
3012304651
Firm city
Worthing
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register