Philips Medical Systems (Cleveland) Inc: Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Recall number
Z-1338-2013
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Status
Terminated
Initiated
2013-03-12
Posted
2013-05-20
Terminated
2015-06-30
Reason
Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.
Root cause
Use error
Action
The firm, Philips Healthcare, sent an "Urgent - Medical Device Correction" letter dated March 15, 2013, to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to disregard unusually higher CTDIvol value for Type-3 DICOM tag 0018:9345 in the PACS system, if seen. The firm is implementing software update versions 4.2.2 or 2.2.2 to correct the issue. A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customers Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Quantity
515 units
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, WV and WY; countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Denmark, France, Germany, Gibraltar, India, Iraq, Ireland, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela.
Lot, serial or model codes
System Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810.  System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741.  System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810.  System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760.  System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549.
510(k) numbers
["K012009"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register