Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage Catheter Kit, Expel" Drainage Catheter APD Drainage Catheter, Expel" Large Capacity Drainage Catheter APD Large Capacity Drainage Catheter.

Recall number
Z-1336-2015
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage Catheter Kit, Expel" Drainage Catheter APD Drainage Catheter, Expel" Large Capacity Drainage Catheter APD Large Capacity Drainage Catheter.
Status
Terminated
Initiated
2015-02-25
Posted
2015-03-27
Terminated
2016-08-12
Reason
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Root cause
Nonconforming Material/Component
Action
Boston Scientific sent an "Urgent Medical Device Correction" letter dated February 25, 2015, to all affected customers. The letter described the problem and the product involved in the recall. The letter informed customers not to use the Expel APD and APDL Drainage catheters for bile drainage and to complete and return the Customer Acknowledgement Form. The letter also noted that the devices are not being retrieved and they are not required to return them to Boston Scientific. For further questions please call (763) 494-7971.
Quantity
918
Distribution
Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.
Lot, serial or model codes
UPN # LOT #  H7493931006200- 17259582, 17362835, 17501623, 17617274. H7493931008200- 17259557, 17297984, 17362834, 17376747. H7493931010200- 17318044, 17376748, 17459140, 17501573, 17617545.  H7493931012250- 17275816, 17501620, 17575123, 17617353. H7493931014250- 17370708, 17604432, 17624928. H7493931106150- 17259581, 17365684, 17617551. H7493931108150- 17244079, 17404400, 17408751, 17493552, 17647113 H7493931108250- 17245000, 17275820, 17297985, 17297986, 17377401, 17416602, 17433865.  H749393110825K1- 17416577. H7493931108450- 17245001, 17408755, 17411332, 17493550. H7493931110150- 17317825, 17518149. H7493931110250- 17228343, 17259558, 17317546, 17338203, 17365681, 17416573, 17438002, 17518150, 17588842, 17588844, 17621339, 17634403, 17634404, 17639037.  H749393111025K1- 17328810, 17403872, 17416575, 17494061. H7493931110450- 17275823, 17416603, 17459143, 17617552. H7493931112250- 17259584, 17275818, 17338205, 17365682, 17601584, 17613485, 17639463.  H7493931112450- 17245002, 17259556, 17365683, 17416604, 17574538, 17617553.  H7493931114250- 17370706, 17588840, 17625218. H7493931114450- 17275817, 17589950, 17624789. H7493932016300- 17328801, 17352479, 17377392, 17408752, 17501571, 17575124, 17612692, 17613484, 17625883, 17626180, 17626181, 17626182.  H7493932018300- 17377395, 17480649, 17614751, 17617350, 17626183. H7493932024400- 17459142, 17492863. H7493932026400- 17328800, 17353160, 17377393, 17617470, 17624927, 17625216, 17625895.
510(k) numbers
["K141335"]
PMA numbers
not on file
Product code
FGE
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register