Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: S T-Bone Plate; M T-Compression Plate; Manufactured and Distributed by Stryker Lebinger GmbH & Co. Freiburg, Germany. The Profyle System is intended for use in internal fixation of the bones of hand and wrist. Examples of these procedures may include but are not limited to replantation, lag screws techniques, joint fusion, corrective osteotomies, and the treatment of fractures

Recall number
Z-1334-2011
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
S T-Bone Plate; M T-Compression Plate; Manufactured and Distributed by Stryker Lebinger GmbH & Co. Freiburg, Germany. The Profyle System is intended for use in internal fixation of the bones of hand and wrist. Examples of these procedures may include but are not limited to replantation, lag screws techniques, joint fusion, corrective osteotomies, and the treatment of fractures
Status
Terminated
Initiated
2006-05-26
Posted
2011-02-16
Terminated
2014-12-08
Reason
A package labeled as a S T-Bone Plate, Catalog Number 57-05260, Lot Code G4M00F1576, may actually contain a M T-Compression Plate, Catalog Number 57-10250, Lot Code G4N00F1584 and vice versa.
Root cause
Process design
Action
Stryker Orthopedics sent an Important Market Withdrawal letter dated May 26, 2006, by Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory, including their hospital consignment location, to identify the product. If the affected product remained in its original packaging, customers were instructed to retrieve all of the affected product lots and return it to their branch or agency warehouse for reconcillation. Reconcile all product on the attached Product Accountability Form and fax a copy of the form to 201-831-6069 within 5 days of receipt of the notice. Customers were asked to return the affected product via Inter-Org Transfer and ship to Product Surveillance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey, USA 07430-2104. Customers were also instructed to contact their Stryker Orthopaedics Customer Service representataive to re-order the product that they were returning. For questions regarding this recall call (201) 831-5825.
Quantity
21
Distribution
Nationwide Distribution
Lot, serial or model codes
S T-Bone Plate, Product number: 57-05260; Lot Code G4M 00F1576; M T-Compression Plate, Product number 57-10260; Lot Code G4N00F1584
510(k) numbers
["K062498"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register