Medtronic Inc, Neurological & Spinal Division: 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N''Vision Clinician Programmer.

Recall number
Z-1334-04
Recalling firm
Medtronic Inc, Neurological & Spinal Division
Product
8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N''Vision Clinician Programmer.
Status
Terminated
Initiated
2004-08-24
Posted
2004-09-23
Terminated
2005-12-13
Reason
With the Model 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N'Vision Clinician Programmer, users may have mistakenly entered a periodic bolus interval into the minutes field, rather than the hours field, resulting in drug overdoses. This issue is limited to programming the SynchroMed and SynchroMed EL pumps.
Root cause
Other
Action
Letters were distributed 8/24/2004 utilizing a courier with signed receipt upon delivery. The letters were signed for upon receipt at the consignee address, and the contents of the letter included a response form which can be faxed or mailed back to Medtronic. If any letters are returned as 'undeliverable', attempts will be made to telephone these consignees. A press release was issued on September 22, 2004 by Medtronic Inc. for all Version AAA 02 Model 8870 software application cards in the U.S.
Quantity
3,218 cards distributed in US, plus 1,248 distributed outside of the US
Distribution
These cards were distributed primarily in the USA, but also to countries in Europe, Middle East and Africa, Canada, Australia, Hong Kong, Latin America and Japan.
Lot, serial or model codes
Software Version AAA 02, BBB 04, BBC 02, and BBD 01
510(k) numbers
not on file
PMA numbers
["P860004"]
Product code
LKK
Firm FEI
2182207
Firm city
Columbia Heights
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register