FDA Medical Device Recalls (openFDA)
Arcoma AB: Omnera 400T Digital Radiographic System
- Recall number
- Z-1331-2022
- Recalling firm
- Arcoma AB
- Product
- Omnera 400T Digital Radiographic System
- Status
- Open, Classified
- Initiated
- 2021-05-05
- Posted
- 2022-07-05
- Terminated
- not on file
- Reason
- There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
- Root cause
- Device Design
- Action
- On May 7, 2021, Arcoma issued a "Urgent Field Safety Notice". The recall notice asked consignees to take the following actions: Identify the device Take note of amendment/reinforcement of Instructions of Use (IFU) and replace all pages in the Cleaning and disinfection chapter with pages in FSN_2021_01_IFU-EN Rev. 1.1. On May 23, 2022, Arcoma issued a Safety Advisory Notice/ Field Safety Notice via E-Mail informing consignees that they would be updating the display handle with the updated version as an additional safety precaution. Arcoma has requested that Canon Medical System, USA ensure that all those who have the affected product, including any others who may need to be informed, receive the safety relevant information provided with this notice and comply with the recommendations herein. A representative of Canon Medical System USA will contact you to schedule a time to replace the display handle with the updated version.
- Quantity
- 71 systems
- Distribution
- US Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA, OR, IN, RI, MI, IL, LA, OH, CO, SC, VA, WA, CA, AZ, NV, AR and the country of Canada.
- Lot, serial or model codes
- UDI/DI:07350008750050; 2001-2003, 2006-2044, 2046-2056, 2118-2126, 2128-2131, 2134-2160, 2164-2175, 2177-2195, 2197-2208, 2210-2212, 2214-2224, 2226-2236.
- 510(k) numbers
- ["K140683"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 3002675731
- Firm city
- Vaxjo
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28