FDA Medical Device Recalls (openFDA)

Arcoma AB: Omnera 400T Digital Radiographic System

Recall number
Z-1331-2022
Recalling firm
Arcoma AB
Product
Omnera 400T Digital Radiographic System
Status
Open, Classified
Initiated
2021-05-05
Posted
2022-07-05
Terminated
not on file
Reason
There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
Root cause
Device Design
Action
On May 7, 2021, Arcoma issued a "Urgent Field Safety Notice". The recall notice asked consignees to take the following actions: Identify the device Take note of amendment/reinforcement of Instructions of Use (IFU) and replace all pages in the Cleaning and disinfection chapter with pages in FSN_2021_01_IFU-EN Rev. 1.1. On May 23, 2022, Arcoma issued a Safety Advisory Notice/ Field Safety Notice via E-Mail informing consignees that they would be updating the display handle with the updated version as an additional safety precaution. Arcoma has requested that Canon Medical System, USA ensure that all those who have the affected product, including any others who may need to be informed, receive the safety relevant information provided with this notice and comply with the recommendations herein. A representative of Canon Medical System USA will contact you to schedule a time to replace the display handle with the updated version.
Quantity
71 systems
Distribution
US Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA, OR, IN, RI, MI, IL, LA, OH, CO, SC, VA, WA, CA, AZ, NV, AR and the country of Canada.
Lot, serial or model codes
UDI/DI:07350008750050; 2001-2003, 2006-2044, 2046-2056, 2118-2126, 2128-2131, 2134-2160, 2164-2175, 2177-2195, 2197-2208, 2210-2212, 2214-2224, 2226-2236.
510(k) numbers
["K140683"]
PMA numbers
not on file
Product code
KPR
Firm FEI
3002675731
Firm city
Vaxjo
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register