Skip to the content

FDA Medical Device Recalls (openFDA)

Cordis US Corp: Vista Brite Tip, Judkins Right 4, REF SM7504, JR3 VBT (Special) 110 cm 6F. Guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.

Recall number
Z-1330-2025
Recalling firm
Cordis US Corp
Product
Vista Brite Tip, Judkins Right 4, REF SM7504, JR3 VBT (Special) 110 cm 6F. Guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Status
Open, Classified
Initiated
2025-02-07
Posted
2025-03-13
Terminated
not on file
Reason
Catheter manufactured at the incorrect length.
Root cause
Employee error
Action
On February 7, 2025 Cordis issued a Urgent Medical Device Recall notification to affected consignees via E-Mail. Cordis asked consignees to take the following actions: 1) Read this Urgent Medical Device Recall letter. 2) Immediately check your inventory for any units from the affected lot in your possession. 3. Identify and quarantine any units. Check all storage and usage locations. 3) Share this letter with others in your facility who need to be made aware of this recall and with any other facility. 4) Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis at the fax number on the form or email to: GMB-CordisFieldAction@cordis.com 5) Return any affected product to the address listed on the form, with reference to your Customer Number which is listed on the form. 6) Maintain awareness of this notice until all affected product has been returned to Cordis and keep a copy of this notice with the affected product.
Quantity
12 units
Distribution
US: WA OUS: Japan
Lot, serial or model codes
UDI: 20705032042991/ Lot Number: 18368326
510(k) numbers
["K021593"]
PMA numbers
not on file
Product code
DQY
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register