FDA Medical Device Recalls (openFDA)

Bio-rad Laboratories, Inc.: Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an unassayed quality control serum to monitor precision of IgM laboratory testing procedures for the analytes listed in this package insert. This product is not intended for use in blood donor screening assays. Analytes claimed with this lot of Liquichek ToRCH Plus IgM Control: Cytomegalovirus (CMV) IgM Herpes Simplex Virus Type 1/2 (HSV-1/2) IgM Rubella Virus IgM Toxoplasma gondii IgM

Recall number
Z-1330-2013
Recalling firm
Bio-rad Laboratories, Inc.
Product
Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an unassayed quality control serum to monitor precision of IgM laboratory testing procedures for the analytes listed in this package insert. This product is not intended for use in blood donor screening assays. Analytes claimed with this lot of Liquichek ToRCH Plus IgM Control: Cytomegalovirus (CMV) IgM Herpes Simplex Virus Type 1/2 (HSV-1/2) IgM Rubella Virus IgM Toxoplasma gondii IgM
Status
Terminated
Initiated
2013-04-15
Posted
2013-05-16
Terminated
2013-08-01
Reason
The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Root cause
Material/Component Contamination
Action
Bio-rad Laboratories initiated this recall on April 18, 2013, via telephone call and certified mailing to customers. The notification letter titled "Important Product Information Notice of Correction" was accompanied by a "Field Correction Notification Acknowledgement" form that notified customers of the product for recall (description and codes), issue (This product contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product), Customers were instructed to immediately discontinue use this lot of Liquichek ToRCH Plus IgM Control. Discard all remaining inventory of this lot of Liquichek ToRCH Plus IgM Control that you have at your location in accordance with the requirements of your local waste management authorities. Handle this product with the same precautions used with patient specimens. In accordance with good laboratory practice, all human source material should be considered potentially infectious. Have the appropriate laboratory personnel complete the acknowledgment form and email or fax it to us. It is very important to us to know that you have received the information provided in this Notice. Contact information (1-800-2-BIORAD)
Quantity
211 boxes and 5 MiniPaks
Distribution
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
Lot, serial or model codes
Lot Number 18972, Catalog Number 229 and 229PX.
510(k) numbers
["K040822"]
PMA numbers
not on file
Product code
JJY
Firm FEI
2016706
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register