Randox Laboratories Ltd.: Liquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that require sample predilution.
- Recall number
- Z-1328-2024
- Recalling firm
- Randox Laboratories Ltd.
- Product
- Liquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that require sample predilution.
- Status
- Open, Classified
- Initiated
- 2024-03-01
- Posted
- 2024-03-20
- Terminated
- not on file
- Reason
- The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.
- Root cause
- Under Investigation by firm
- Action
- Consignees were sent a MEDICAL DEVICE CORRECTION NOTIFICATION dated 03/06/2024. The notification instructs consignees to review results generated with the affected batches in line with the clinical profile of the patient, discuss the contents of the provided recall notice with the facility's Medical Director, and complete and return the response form to technical.services@randox.com within five working days. The notice should be shared with all customers if product was further distributed and to those who need to be aware within consignee organizations. Consignees with questions should reach out to Randox Technical Services.
- Quantity
- 1,764 units
- Distribution
- Domestic: CA, FL, IL, ME, NJ, OH, & OR.
- Lot, serial or model codes
- Catalog No. IT2691; GTIN 05055273204032; Batch No./Lot No. (Exp. Date): 647343/2154IT-2158IT (03/28/25), 647342/2154IT-2158IT (03/28/25), 634886/2112IT-2116IT (07/28/24), 634887/2112IT-2116IT (07/28/24), 627224/2112IT-2116IT (07/28/24), 627222/2112IT-2116IT (07/28/24)
- 510(k) numbers
- ["K061056"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28