Skip to the content

FDA Medical Device Recalls (openFDA)

Nova Biomedical Corporation: Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Ca++, total hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), reduced hemoglobin (HHb), Oxygen content (O2Ct) and Oxygen capacity (O2Cap) in heparinized whole blood; Na+, K+, CL-; Ca++, Mg++, Glucose heparinized whole blood, serum, or plasma.

Recall number
Z-1328-2009
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Ca++, total hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), reduced hemoglobin (HHb), Oxygen content (O2Ct) and Oxygen capacity (O2Cap) in heparinized whole blood; Na+, K+, CL-; Ca++, Mg++, Glucose heparinized whole blood, serum, or plasma.
Status
Terminated
Initiated
2008-08-11
Posted
2009-05-07
Terminated
2011-07-22
Reason
Erroneous, low creatinine results on CCX analyzer
Root cause
Process control
Action
Initial: USA and Canada - Customer Advisory Notice notification by phone contact with key customer contact facilitated using a Field Correction Script. This contact will be documented within the Nova Biomedical Technical Support database (Pivotal) and on a Field Correction Checklist. All affected inventory will be replaced at customer sites. International - Customer Advisory Notice and Faxback Form being communicated and sent to Nova Biomedical Subsidiaries and Distributors by three methods of delivery (Email, Fax and Phone Call). Follow-up: USA and Canada - Sending the Customer Advisory Notice by email and/or fax to the key customer contact from the initial phone contact. International - Sending the Customer Advisory Notice by email and/or fax to the key contact.
Quantity
1230
Distribution
Worldwide Distribution -- USA, Canada, Germany, France, Czech Republic, Taiwan, Japan, Thailand, Italy, Hungary, India, UK, Ireland, Venezuela, Brazil, and Hong Kong.
Lot, serial or model codes
Lots 806050, 806182, 806242, 806567, 806721, 806752, and 807246.
510(k) numbers
["K022746"]
PMA numbers
not on file
Product code
CGL
Firm FEI
1219029
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register