Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.

Recall number
Z-1327-2016
Recalling firm
Synthes (USA) Products LLC
Product
Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Status
Terminated
Initiated
2016-02-23
Posted
2016-04-05
Terminated
2016-09-08
Reason
Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
Root cause
Nonconforming Material/Component
Action
DePuy Synthes sent an Urgent Notice Medical Device Recall letter dated February 23, 2016 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were notified by mail of a potential patient impact. They were instructed to review existing inventories and call DePuy Synthes at 1800-479-6329 to obtain a Return Authorization (RA) Number and replacement order. After completing the verification section on page 4 of the letter, they were advised to return the verification section (page 3) with the product to Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132. A copy of the completed verification section is to sent to DePuy Synthes by: Fax 877-931-0643 or scan/email Synthes8097@stericycle.com.
Quantity
150 units distributed in US, 1439 units distributed OUS (foreign)
Distribution
Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.
Lot, serial or model codes
Titanium Low Profile Neuro Screw, Self-Drilling, 3mm, part number # 400.833E, Lot numbers: 9958565, 9951621, 9952976, 9955377, 9962314, 9968868, 9969645
510(k) numbers
["K031807"]
PMA numbers
not on file
Product code
DZL
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register