FDA Medical Device Recalls (openFDA)
Smiths Medical PM, Inc.: Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model 9200, catalog numbers: 920654225, 920674225, 920674235, 925454220, 925454235, 925454320, 925454325, 925454330, 925454335, 925457325, 925474325, 925654225, 925654325, 925754325, 925754335, and 92M774325. Intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of a hospital or clinic where low-end monitoring systems are needed.
- Recall number
- Z-1326-2009
- Recalling firm
- Smiths Medical PM, Inc.
- Product
- Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model 9200, catalog numbers: 920654225, 920674225, 920674235, 925454220, 925454235, 925454320, 925454325, 925454330, 925454335, 925457325, 925474325, 925654225, 925654325, 925754325, 925754335, and 92M774325. Intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of a hospital or clinic where low-end monitoring systems are needed.
- Status
- Terminated
- Initiated
- 2008-12-22
- Posted
- 2009-05-15
- Terminated
- 2011-12-17
- Reason
- Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
- Root cause
- Device Design
- Action
- An Urgent Product Safety Information, Voluntary Recall Notification, dated 12/22/08, was sent to affected consignees. The notification contained a Safety Action Bulletin which provided actions to be taken. The Safety Action Bulletin described the issue, requested customer to contact Smiths Medical to ensure proper mounting hardware is utilized for those consignees using a roll stand or wall mount. If the consignee further distributed the product, they were asked to forward on the Safety Action Bulletin to the end user. Page 3 of the Safety Action Bulletin was asked to be completed and returned to Smiths Medical PM, Inc. A second "Urgent Product safety Information Voluntary Recall Notification" letter dated January 16, 2009 was sent to consignees. The second letter included a risk statement associated with the recalled product.
- Quantity
- 143 monitors (17 US, 126 OUS)
- Distribution
- Worldwide Distribution -- USA including states of FL, GA, NJ, NY, and TN and countries of Argentina, Brazil, Ecuador, Greece, Peru, Romania, Russia, Singapore, Slovenia, Trinidad and Tobago, United Kingdom, and Uruguay.
- Lot, serial or model codes
- serial numbers: AM08070014, AM08070015, AM08080004, AM08080005, AM08080006, AM08080007, AM08080008, AM08080009, AM08080010, AM08080011, AM08080012, AM08080013, AM08080014, AM08080015, AM08080016, AM08080017, AM08080018, AM08080019, AM08080032, AM08080033, AM08080034, AM08080035, AM08080036, AM08080037, AM08080038, AM08080039, AM08080040, AM08080041, AM08080042, AM08080043, AM08080044, AM08080045, AM08080046, AM08080047, AM08080048, AM08080049, AM08080050, AM08080051, AM08080052, AM08080075, AM08080076, AM08080077, AM08080078, AM08080079, AM08090058, AM08090059, AM08090060, AM08090061, AM08090069, AM08090070, AM08090071, AM08090072, AM08090073, AM08090074, AM08090075, AM08090076, AM08090077, AM08090078, AM08090079, AM08090080, AM08090081, AM08090082, AM08090083, AM08090084, AM08090085, AM08090086, AM08090092, AM08090093, AM08090094, AM08090095, AM08090096, AM08090097, AM08100000, AM08100001, AM08100002, AM08100003, AM08100004, AM08100005, AM08100006, AM08100007, AM08100008, AM08100042, AM08100043, AM08100045, AM08100051, AM08100052, AM08100053, AM08100054, AM08100056, AM08100075, AM08100076, AM08100077, AM08100078, AM08100080, AM08100082, AM08100083, AM08100085, AM08100086, AM08100087, AM08100088, AM08100089, AM08100091, AM08100092, AM08100093, AM08100094, AM08100095, AM08100096, AM08100097, AM08100098, AM08100099, AM08100100, AM08100101, AM08100146, AM08110000, AM08110001, AM08110003, AM08110004, AM08110005, AM08110008, AM08110009, AM08110010, AM08110011, AM08110012, AM08110013, AM08110014, AM08110015, AM08110016, AM08110018, AM08110019, AM08110026, AM08110027, AM08110028, AM08110029, AM08110030, AM08110031, AM08110034, AM08110035, AM08110036, AM08110037, AM08110054, AM08110055, AM08110056, and AM08110058.
- 510(k) numbers
- ["K010770","K011177","K030407","K982279"]
- PMA numbers
- not on file
- Product code
- CCK
- Firm FEI
- 2183502
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28