FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A

Recall number
Z-1325-2025
Recalling firm
Intuitive Surgical, Inc.
Product
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A
Status
Open, Classified
Initiated
2025-02-26
Posted
2025-03-12
Terminated
not on file
Reason
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
Root cause
Device Design
Action
On February 26, 2025, Intuitive Surgical issued a Urgent: Medical Device Correction to affected consignees via E-Mail. Intuitive asked consignees to take the following actions: 1. Read and understand this notification. 2. Share this notification with all personnel at your site who perform and support da Vinci procedures. 3. Inform affected personnel when the correction has been completed. 4. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. 6. Please inform Intuitive of any Adverse Events* or quality problems concerning the use of the subject devices via the standard complaint process 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Complete and submit the report Online Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. An Intuitive representative will schedule a visit to replace the affected foot tray assembly.
Quantity
439 units
Distribution
Worldwide distribution - US Nationwide and the countries of Alaska, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the country of South Korea.
Lot, serial or model codes
UDI: 00886874119747/ All da Vinci dV5 consoles and foot trays assemblies manufactured prior to 12/20/2024.
510(k) numbers
["K232610"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register