PENTAX of America Inc: Pentax Medical C2 CryoBalloon Standard Focal Catheter, Model FG 1028, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.

Recall number
Z-1325-2019
Recalling firm
PENTAX of America Inc
Product
Pentax Medical C2 CryoBalloon Standard Focal Catheter, Model FG 1028, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.
Status
Terminated
Initiated
2018-12-03
Posted
not on file
Terminated
2020-08-14
Reason
Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.
Root cause
Under Investigation by firm
Action
Letters dated 12/3/2018 were issued via USPS mailing on 12/3/2018 explaining the issue and instructing the customer to discontinue use of the device and to ensure all potential users in the customer facility are aware of the recall notification and recommended actions. For questions regarding this action, please contact PENTAX Medical Customer Service. Tel: 650-318-5899 (8:30 AM 5:00 PM, Monday Friday, PST) Fax: 650-318-5899 Email: RMA@C2Therapeutics.com
Quantity
157 devices
Distribution
Distribution was made to CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada, Germany, and the Netherlands.
Lot, serial or model codes
Lot/Serial numbers: 06152018-02, 06252018-03, 07062018-01, 07232018-02, 08102018-01, 08172018-06, 08292018-04, and 9172018-01.
510(k) numbers
["K163684"]
PMA numbers
not on file
Product code
GEH
Firm FEI
3008780134
Firm city
Redwood City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register