Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Trabecular Metal Technology, Inc.: NexGen TRABECULAR METAL FEMORAL AUGMENT BLOCK DISTAL ONLY, 20 MM Augment, Size D, Sterile, Rx Only. Product Usage: Intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution-Rotating Hinge Knee (RHK) System, the Trabecular Metal Augments are for cemented use only. When used with the NexGen Complete Knee Solution-Legacy Constrained Condylar Knee System, the Trabecular Metal Augments are for cementless or cemented use.

Recall number
Z-1325-2015
Recalling firm
Zimmer Trabecular Metal Technology, Inc.
Product
NexGen TRABECULAR METAL FEMORAL AUGMENT BLOCK DISTAL ONLY, 20 MM Augment, Size D, Sterile, Rx Only. Product Usage: Intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution-Rotating Hinge Knee (RHK) System, the Trabecular Metal Augments are for cemented use only. When used with the NexGen Complete Knee Solution-Legacy Constrained Condylar Knee System, the Trabecular Metal Augments are for cementless or cemented use.
Status
Terminated
Initiated
2015-02-04
Posted
2015-03-26
Terminated
2015-07-16
Reason
One unit of the 20 mm augments was mislabeled as 5 mm and it was subsequently distributed.
Root cause
Error in labeling
Action
Zimmer Trabecular Metal Technology sent an email dated 12/10/2014 to their sole customer. The customer was requested to return the affected product.
Quantity
1 unit
Distribution
US Distribution in the state of CA
Lot, serial or model codes
20 MM - Part No. 00-5490-034-24, Lot #613574282T 5 MM - Part No. 00549003410, Lot #61357484T
510(k) numbers
["K040487"]
PMA numbers
not on file
Product code
MBH
Firm FEI
3005751028
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register