Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Renal Therapies Group, LLC: Fresenius Hemoflow F3 Dialyzer Part Number: 0520165A Indicated for Acute and Chronic Hemodialysis

Recall number
Z-1324-2016
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Hemoflow F3 Dialyzer Part Number: 0520165A Indicated for Acute and Chronic Hemodialysis
Status
Terminated
Initiated
2016-02-26
Posted
2016-04-05
Terminated
2017-06-26
Reason
Dialyzer header leak due to possible improper torque
Root cause
Process control
Action
The firm, Fresenius Medical Care Renal Therapies Group, LLC issued an "URGENT MEDICAL DEVICE RECALL" letter dated 2/24/16 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of your Hemoflow F3 Dialyzer; place all affected units in a secure, segregated area; contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product, and fill out and return the attached Reply Form via fax to: 781-699-9796 ATTN: Quality Department or scan and email to: NOTIFYRA@fmc-na.com at your earliest convenience. If you have any additional medical concerns or questions, please contact Medical Information and Communication at 855-616-2309 or Website: www.fresenius-medinfo.com.
Quantity
1,650 units
Distribution
US Nationwide Distribution including states of: AZ, CA, CT, DC, GA, IL, IN, MI, MN, MO, MS, NY, OH, PA, SC, SD, TX, and WA.
Lot, serial or model codes
Lot Number: 15KU04011 Exp. Date : 08/31/2013
510(k) numbers
["K874872"]
PMA numbers
not on file
Product code
FJI
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register