Skip to the content

FDA Medical Device Recalls (openFDA)

Nobel Biocare Usa Llc: Implant Retrieval Kit - Product Usage: The kit itself does not have a specific intended Use, but included devices have specific intended use as shown above.

Recall number
Z-1323-2020
Recalling firm
Nobel Biocare Usa Llc
Product
Implant Retrieval Kit - Product Usage: The kit itself does not have a specific intended Use, but included devices have specific intended use as shown above.
Status
Terminated
Initiated
2019-03-14
Posted
2020-02-19
Terminated
2024-02-15
Reason
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Root cause
Packaging process control
Action
On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instructions provided in this Recall Letter and return the attached Customer Reply Form to the firm via email or fax at 714- 282-5073.
Quantity
87 units
Distribution
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
Lot, serial or model codes
857717 (01)07332747083750(10)857717 28-Dec-2018 857597 (01)07332747083750(10)857597 31-Aug-2018 857656 (01)07332747083750(10)857656 22-Oct-2018 818712 (01)07332747083750(10)818712 9-Nov-2018 818747 (01)07332747083750(10)818747 9-Nov-2018 818647 (01)07332747083750(10)818647 10-Oct-2018 857578 (01)07332747083750(10)857578 23-Aug-2018 857610 (01)07332747083750(10)857610 10-Sep-2018 857615 (01)07332747083750(10)857615 13-Sep-2018 857653 (01)07332747083750(10)857653 10-Oct-2018 857677 (01)07332747083750(10)857677 7-Nov-2018 857694 (01)07332747083750(10)857694 3-Dec-2018 857708 (01)07332747083750(10)857708 14-Dec-2018 857711 (01)07332747083750(10)857711 20-Dec-2018 857729 (01)07332747083750(10)857729 8-Jan-2019 857732 (01)07332747083750(10)857732 15-Jan-2019 818682 (01)07332747083750(10)818682 5-Nov-2018 818811 (01)07332747083750(10)818811 18-Dec-2018 Catalog# 37470
510(k) numbers
["K990846"]
PMA numbers
not on file
Product code
DZI
Firm FEI
1000125598
Firm city
Yorba Linda
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.