Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Weck Vista Universal Laparoscopic Port, Size 5/10mm x 100mm, Catalog No. 405910, Weck Vista Universal Laparoscopic Port, Size 5/10/12mm x 100mm, Catalog No. 405912; Weck Vista Universal Laparoscopic Port, Size 5/10/12mm x 125mm, Catalog No. 40591213. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals.

Recall number
Z-1323-2014
Recalling firm
Teleflex Medical
Product
Weck Vista Universal Laparoscopic Port, Size 5/10mm x 100mm, Catalog No. 405910, Weck Vista Universal Laparoscopic Port, Size 5/10/12mm x 100mm, Catalog No. 405912; Weck Vista Universal Laparoscopic Port, Size 5/10/12mm x 125mm, Catalog No. 40591213. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals.
Status
Terminated
Initiated
2014-03-11
Posted
2014-04-02
Terminated
2016-01-28
Reason
Complaints of leakage of insufflation gas through the device.
Root cause
Nonconforming Material/Component
Action
Letters were issued on March 11, 2014 to customers informing them of the recall. The letter asked the consignee to immediately discontinue use and quarantine any products on hand. The letter included a Recall Acknowledgement Form which is to be returned. A letter was also issued to distributors on March 11, 2014, requesting them to conduct a sub-recall.
Quantity
55,521 total
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, MD, MA, MI, MN, MS, MO, NH, NJ, NY, NC, ND, OH, OR, PA, SC, TN, UT, VA, WA, WV, WI and the countries of Australia, Brazil, Canada, China, Israel, Korea, Malaysia, Singapore, and Vietnam.
Lot, serial or model codes
Lots Catalog No. 405910 - 01G1200242, 01G1200242, 01G1200341, 01J1200100, 01J1200100, 01J1200378, 01J1200461, 01K1200266, 01K1200171, 01K1200266, 01L1200182, 01A1300139, 01A1300139, 01B1300050, 01D1300318, 01E1300139, 01E1300217, 01E1300432; Catalog No. 405912 01J1200462, 01K1200129, 01K1200377, 01K1200596, 01K1200611, 01K1200611, 01L1200369, 01L1200369, 01L1200369, 01A1300106, 01A1300106, 01A1300106, 01E1300218, 01E1300311, 01F1300135, 01F1300079; Catalog No. 40591213- 01J1200416, 01K1200612, 01L1200160, 01A1300418, 01A1300427, 01E1300056, 01E1300052, 01E1300055
510(k) numbers
["K082156"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
3005747797
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register