Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH 401222RH 401223RH 401226RH 401227RH 401260RH 401261RH 401306RH 401309RH 401312RH 401430RH 401433RH 401435RH 401441RH 401442RH 401443RH 401448RH 401449RH 401450RH 401451RH 401474RH 401860RH 401877RH 401878RH 401890RH 401891RH 401994RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5-5-5mm 402012RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5mm 402004RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 2-5-2mm 402010RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 5-5-5mm 402008RH 402009RH

Recall number
Z-1322-2026
Recalling firm
Medline Industries, LP
Product
Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH 401222RH 401223RH 401226RH 401227RH 401260RH 401261RH 401306RH 401309RH 401312RH 401430RH 401433RH 401435RH 401441RH 401442RH 401443RH 401448RH 401449RH 401450RH 401451RH 401474RH 401860RH 401877RH 401878RH 401890RH 401891RH 401994RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5-5-5mm 402012RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5mm 402004RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 2-5-2mm 402010RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 5-5-5mm 402008RH 402009RH
Status
Open, Classified
Initiated
2025-12-22
Posted
2026-02-06
Terminated
not on file
Reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Root cause
Under Investigation by firm
Action
On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
Quantity
48
Distribution
Distribution US nationwide.
Lot, serial or model codes
401211RH UDI-DI 10197344043736 Lot EP250317; 401222RH UDI-DI 10197344043712 Lot EP250317; 401223RH UDI-DI 10197344043705 Lot EP250317; 401226RH UDI-DI 10197344043699 Lot EP250307 EP250317; 401227RH UDI-DI 10197344043682 Lots EP250307 EP250317; 401260RH UDI-DI 10197344043668 Lot EP250317; 401261RH UDI-DI 10197344043651 Lot EP250317; 401306RH UDI-DI 10197344043620¿ Lot EP250317; 401309RH UDI-DI 10197344043613 Lot EP250317; 401312RH UDI-DI 10197344043583¿ Lot EP250307; 401430RH UDI-DI 10197344043491 Lots EP250307 EP250317; 401433RH UDI-DI 10197344043484 Lots EP250307 EP250317; 401435RH UDI-DI 10197344043460 Lots EP250307 EP250317; 401441RH UDI-DI 10197344043439 Lots EP250307 EP250317; 401442RH UDI-DI 10197344044245 Lot EP250307; 401443RH UDI-DI 10197344044252 Lots EP250307 EP250317; 401448RH UDI-DI 10197344044283 Lot EP250307; 401449RH UDI-DI 10197344044290 Lot EP250307; 401450RH UDI-DI 10197344044306 Lots EP250307 EP250317; 401451RH UDI-DI 10197344044313 EP250307 EP250317; 401474RH UDI-DI 10197344044351 Lot EP250307; 401860RH UDI-DI 10197344044382 Lots EP250307 EP250317; 401877RH UDI-DI 10197344044436 Lots EP250307 EP250317; 401878RH UDI-DI 10197344044443 Lot EP250307; 401890RH UDI-DI 10197344044450 Lots EP250307 EP250317; 401891RH UDI-DI 10197344044467 Lot EP250307; 401994RH UDI-DI 10197344044542 Lot EP250317; 402012RH UDI-DI 10197344043835 Lots EP250307 EP250317; 402004RH UDI-DI 10197344044566 Lots EP250307 EP250317; 402010RH UDI-DI 10197344044597 Lots EP250307 EP250317; 402008RH UDI-DI 10197344044573 Lot EP250317; 402009RH UDI-DI 10197344044580 Lots EP250307 EP250317 ***Updated 5/18/26*** LOT EP250203 added to the scope of action for the following SKUs 401222RH 401223RH 401226RH 401227RH 401261RH 401430RH 401433RH 401441RH 401442RH 401443RH 401448RH 401449RH 401450RH 401451RH 401860RH 401891RH 402004RH 402010RH 402012RH
510(k) numbers
["K241156"]
PMA numbers
not on file
Product code
NLH
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register