Skip to the content

FDA Medical Device Recalls (openFDA)

Clinical Diagnostic Systems: VITROS Immunodiagnostic Products TSH Reagent Pack

Recall number
Z-1322-2018
Recalling firm
Clinical Diagnostic Systems
Product
VITROS Immunodiagnostic Products TSH Reagent Pack
Status
Terminated
Initiated
2018-01-11
Posted
2018-04-10
Terminated
2020-09-08
Reason
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
Root cause
Nonconforming Material/Component
Action
Distributors instructed to disgard remaining inventory, complete Confirmation of Receipt form, and firm promised to replace product. Distributors instructed to send letter to customers.
Quantity
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)
Distribution
Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
Lot, serial or model codes
Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017.
510(k) numbers
["K083173"]
PMA numbers
not on file
Product code
JLW
Firm FEI
3003939841
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register