Skip to the content

FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Proplege Coronary Sinus Catheter model PR9, 11 French Introducer, kitted with PR9. Product Usage: The catheter introducer sheath kit provides access to the central venous circulation to facilitate catheter and guidewire insertion. The Proplege coronary sinus catheter provides retrograde cardioplegia during cardiopulmonary by-pass surgery.

Recall number
Z-1322-2013
Recalling firm
Edwards Lifesciences, LLC
Product
Proplege Coronary Sinus Catheter model PR9, 11 French Introducer, kitted with PR9. Product Usage: The catheter introducer sheath kit provides access to the central venous circulation to facilitate catheter and guidewire insertion. The Proplege coronary sinus catheter provides retrograde cardioplegia during cardiopulmonary by-pass surgery.
Status
Terminated
Initiated
2013-01-25
Posted
2013-05-14
Terminated
2017-02-16
Reason
Edwards Lifesciences is recalling certain Proplege Coronary Sinus Catheter, model PR9 due to confirmed reports of blood leakage through the hemostasis valve in the introducer that is kitted with the catheter.
Root cause
Device Design
Action
Edwards Lifesciences sent an Urgent Field Safety Notice letter dated February 1, 2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review entire inventory for the affected product, quarantine and return to Edwards. Contact Customer Services at 800-424-3278 for any question, to obtain a Returned Goods Authorization number and replacement product.
Quantity
807
Distribution
Worldwide Distribution - USA Nationwide including the states of: MD, NY, PA, CA, FL, AZ, AL, TX, VA, OH, MO, NC, GA, TN, CO, OR, SC, ME, IL, HI, NJ, WA, WI, CT, KY, LA and Canada.
Lot, serial or model codes
Model PR9, lot numbers: 59385159, 59385164, 59385204, 59398000, 59398006, 59398008, 59400402, 59423199, 59423200, 59385159, 59385164, 59385204, 59398000, 59398006.
510(k) numbers
["K120780"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1713910
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register