Skip to the content

FDA Medical Device Recalls (openFDA)

ArthroCare Corporation: TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01

Recall number
Z-1321-2016
Recalling firm
ArthroCare Corporation
Product
TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01
Status
Terminated
Initiated
2016-03-03
Posted
2016-04-05
Terminated
2017-05-17
Reason
During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.
Root cause
Package design/selection
Action
Consignees were notified via letter on 3/3/2016. Consignees were instructed to quarantine the recalled products and complete an Inventory Return Certification Form.
Quantity
26,903 devices
Distribution
Worldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.
Lot, serial or model codes
1031506, 1031507, 1032123, 1032325, 1032741, 1032742, 1033443, 1033965, 1033966, 1034397, 1039120, 1041765, 1044089, 1044090, 1045862, 1048237, 1048933, 1049185, 1052071, 1056212, 1060586, 1060979, 1063394, 1064599, 1067696, 1067748, 1069174, 1071358, 1074125, 1075714, 1070521, 1070812, 1071974, 1078025, 1077499, 1082811, 1084095, 1084094, 1085458, 1086436, 1086437, 1087917, 1094313, 1097573, 1103824, 1107183, 1107297, 1113562, 1113563, 1114534, 1114535, 1114536, 1115896, 1031518, 1032308, 1032309, 1032310, 1032755, 1032756, 1033972, 1033973, 1034225, 1034226, 1035283, 1035284, 1038547, 1043507, 1046556, 1047064, 1048722, 1049266, 1051812, 1052070, 1054443, 1055730, 1059865, 1063661, 1067691, 1067273, 1067747, 1068019, 1069172, 1071357, 1072853, 1072854, 1070497, 1070808, 1071971, 1076849, 1077450, 1079140, 1079220, 1078115, 1081753, 1082361, 1082810, 1086380, 1094312, 1094312, 1094721, 1097571, 1101041, 1107295, 1109045, 1111935, 1111936, 1113534, 1113535, 1113536, 1114532, 1115888, 1115891, 1116959, 1116130, 1116720, 1117209, 1117214, 1031341, 1032311, 1033761, 1034446, 1035691, 1035692, 1041775, 1043962, 1046557, 1048236, 1050067, 1052651, 1057008, 1058808, 1059867, 1063665, 1064594, 1067695, 1068021, 1071319, 1071320, 1073441, 1074876, 1077498, 1080520, 1114533, 1115893, 1031340, 1032962, 1033266, 1045359, 1047065, 1051331, 1056849, 1064592, 1067694, 1070709
510(k) numbers
["K080282"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3010813691
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register