Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH

Recall number
Z-1320-2026
Recalling firm
Medline Industries, LP
Product
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
Status
Open, Classified
Initiated
2025-12-22
Posted
2026-02-06
Terminated
not on file
Reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Root cause
Under Investigation by firm
Action
On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
Quantity
311
Distribution
Distribution US nationwide.
Lot, serial or model codes
81105RH UDI-DI 10197344133253 Lot EP250618; 81172RH UDI-DI 10197344133284 Lot EP250618; 81202RH UDI-DI 10197344133307 Lot EP250618; 81483RH UDI-DI 10197344133444 Lot EP250618; 81102RH UDI-DI (case) 20197344133236¿ (ea) 10197344133239 Lot EP250618; 81104RH UDI-DI 10197344133246 Lot EP250618; 81174RH UDI-DI 10197344133291 Lot EP250618; 81531RH UDI-DI 10197344133512 Lot EP250618; 81532RH UDI-DI 10197344133529 Lot EP250618; 81107RH UDI-DI 10197344133260 Lot EP250618; 81402RH UDI-DI 10197344133352 Lot EP250618; 81404RH UDI-DI 10197344133376 Lot EP250618; 81405RH UDI-DI 10197344133383 Lot EP250618; 81474RH UDI-DI 10197344133437 Lot EP250618; 81473RH UDI-DI 10197344133420 Lot EP250618
510(k) numbers
["K242225"]
PMA numbers
not on file
Product code
NLH
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register