FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
- Recall number
- Z-1320-2026
- Recalling firm
- Medline Industries, LP
- Product
- Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
- Status
- Open, Classified
- Initiated
- 2025-12-22
- Posted
- 2026-02-06
- Terminated
- not on file
- Reason
- These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
- Root cause
- Under Investigation by firm
- Action
- On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
- Quantity
- 311
- Distribution
- Distribution US nationwide.
- Lot, serial or model codes
- 81105RH UDI-DI 10197344133253 Lot EP250618; 81172RH UDI-DI 10197344133284 Lot EP250618; 81202RH UDI-DI 10197344133307 Lot EP250618; 81483RH UDI-DI 10197344133444 Lot EP250618; 81102RH UDI-DI (case) 20197344133236¿ (ea) 10197344133239 Lot EP250618; 81104RH UDI-DI 10197344133246 Lot EP250618; 81174RH UDI-DI 10197344133291 Lot EP250618; 81531RH UDI-DI 10197344133512 Lot EP250618; 81532RH UDI-DI 10197344133529 Lot EP250618; 81107RH UDI-DI 10197344133260 Lot EP250618; 81402RH UDI-DI 10197344133352 Lot EP250618; 81404RH UDI-DI 10197344133376 Lot EP250618; 81405RH UDI-DI 10197344133383 Lot EP250618; 81474RH UDI-DI 10197344133437 Lot EP250618; 81473RH UDI-DI 10197344133420 Lot EP250618
- 510(k) numbers
- ["K242225"]
- PMA numbers
- not on file
- Product code
- NLH
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28