Beckman Coulter Inc.: Access Immunoassay Systems Accu Tnl Reagent Kit when used in conjunction with UniCel Dxl Access Immunoassay Systems; Reagent Part Numbers: A78803 & 33340; Calibrator Part Number: 33345 The intended use: The Access AccuTnI assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac Troponin I determination also aids in the risk stratification of patients with unstable angina or non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction, or increased probability of ischemic events requiring urgent revascularization procedures.
- Recall number
- Z-1318-2010
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access Immunoassay Systems Accu Tnl Reagent Kit when used in conjunction with UniCel Dxl Access Immunoassay Systems; Reagent Part Numbers: A78803 & 33340; Calibrator Part Number: 33345 The intended use: The Access AccuTnI assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac Troponin I determination also aids in the risk stratification of patients with unstable angina or non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction, or increased probability of ischemic events requiring urgent revascularization procedures.
- Status
- Terminated
- Initiated
- 2010-02-06
- Posted
- 2010-04-12
- Terminated
- 2012-05-09
- Reason
- The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with UniCel Dxl systems have been demonstrated to have a positive bias compared to values obtained with Access or Access 2 systems.
- Root cause
- Other
- Action
- An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM. If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
- Quantity
- 196,071 kits
- Distribution
- Worldwide distribution: USA and Canada
- Lot, serial or model codes
- Lot Numbers: 716707, 720541, 720542, 721451, 721452, 722677, 722678, 821024, 821025, 821515, 821663, 823314, 823523, 823524, 823850, 825783, 825784, 826258, 826259, 827876, 828479, 829974, 831218, 850000, 850003, 850004, 850005, 850010, 850011, 850013, 908322, 908323, 908917, 908918, 909803, 909804, 911338, 911339, 911748, 912628, 912629, 913131, 913132, 913265, 913274, 913500, 913821, 914632, 915132, 915417, 915670, 915850, 916720, 916721, 916721A, 917611, 918460 Note: These Lot Numbers represent all reagent lots released to field that could be used in conjunction with Unicel DxI Access Immunoassay Systems.
- 510(k) numbers
- ["K010429","K021814"]
- PMA numbers
- not on file
- Product code
- MMI
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28