Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Recall number
Z-1317-2018
Recalling firm
Smith & Nephew, Inc.
Product
D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R
Status
Terminated
Initiated
2018-03-07
Posted
2018-03-20
Terminated
2018-10-31
Reason
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
Root cause
Under Investigation by firm
Action
The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.
Quantity
59 units
Distribution
US, Puerto Rico, Malaysia
Lot, serial or model codes
Lot Number 17GM04522
510(k) numbers
K132296
PMA numbers
not on file
Product code
HWC
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source