FDA Medical Device Recalls (openFDA)
Hill-Rom, Inc.: Hill-Rom TotalCare Bed, Model P1900. The TotalCare (R) Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare (R) Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post
- Recall number
- Z-1317-2009
- Recalling firm
- Hill-Rom, Inc.
- Product
- Hill-Rom TotalCare Bed, Model P1900. The TotalCare (R) Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare (R) Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post
- Status
- Terminated
- Initiated
- 2009-03-09
- Posted
- 2009-05-01
- Terminated
- 2010-10-13
- Reason
- An e-ring may not have been installed during production, which could result in an inability to latch the siderail.
- Root cause
- Process control
- Action
- Consignees were notified by an Urgent Medical Device Correction letter dated 3/9/09 and were instructed that a Hill-Rom technician will visit each consignee to inspect and to correct the beds, if necessary. The firm asks that consignees locate and test their beds to make sure the siderails are latching correctly; forward a copy of the letter to users that were loaned or purchased the device and notify Hill-Rom of their new locations or owners; and if affected beds were disposed of or are no longer in use, to inform the firm of their serial numbers. If there are questions customers should contact their Hill-Rom Technical Support for their geographic region.
- Quantity
- 53 beds
- Distribution
- Worldwide Distribution -- USA and Canada.
- Lot, serial or model codes
- Serial numbers from J329AM1149 through J339AM1343.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 1824206
- Firm city
- Batesville
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28