Ethicon, Inc.: Orthocord Suture, DePuy Mitek, a Johnson & Johnson Company; . DePuy Mitek, Inc. 325 Paramount Drive, Raynham, MA 02767 Sutures are intended for general soft tissue approximation and/or ligation.

Recall number
Z-1316-2009
Recalling firm
Ethicon, Inc.
Product
Orthocord Suture, DePuy Mitek, a Johnson & Johnson Company; . DePuy Mitek, Inc. 325 Paramount Drive, Raynham, MA 02767 Sutures are intended for general soft tissue approximation and/or ligation.
Status
Terminated
Initiated
2009-02-02
Posted
2009-05-12
Terminated
2009-10-21
Reason
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
Root cause
Process control
Action
Urgent Voluntary Product Recall notification letters were sent via UPS Next Day Air with receipt on February 2, 2009. Letters were sent to customers, Directors of Surgical Services, Director of Risk Management, Materials Manager and Distributors. The letters requested that customers immediately discontinue use of the products with lot numbers listed. Customers are to identify and return only the boxes identified in the attached tables. The affected product should be returned to Ethicon via the pre-paid authorized shipping label. Customers that are not returning affected product should mail the enclosed business reply card to Ethicon. Customers should direct questions regarding this recall to 1-888-202-3694 or their Ethicon sales representative. 5/13/09: It was discovered that a lot of PDS II sutures that had been previously identified as affected but not distributed, were mistakenly distributed. Revised letters including this number were sent to 25 customers on 5/13/09.
Quantity
1,758,360 total
Distribution
Worldwide Distribution -- USA, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Greece, Hong Kong, Israel, Japan, Korea, Luxemburg, Mexico, New Zealand, Norway, Panama, Puerto Rico, Sweden, Taiwan, United Kingdom, Uruguay, and Venezuela.
Lot, serial or model codes
Lot # Expiration Date AP2702 2011-11-30 12:00AM AP2546 2011-11-30 12:00AM
510(k) numbers
["K040004"]
PMA numbers
not on file
Product code
NEW
Firm FEI
2210968
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register