Stryker Sustainability Solutions: Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-235; Color Cuff Single Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-135; Color Cuff Dual Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-235;

Recall number
Z-1315-2023
Recalling firm
Stryker Sustainability Solutions
Product
Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-235; Color Cuff Single Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-135; Color Cuff Dual Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-235;
Status
Open, Classified
Initiated
2023-03-01
Posted
2023-04-01
Terminated
not on file
Reason
Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.
Root cause
Process control
Action
On 3/1/23, recall notices started being distributed to customers. Customers were asked to do the following: 1) Follow the IFU and inspect for tears. Do not use the device if any damage is noted. Return the device and packaging to the recalling firm, if it is not in acceptable condition for surgery. 2) To avoid a pinched or unseated O-ring upon connection of the tourniquet cuff to the tourniquet controller, hold down the connector release on the tourniquet controller and take extra caution to ensure connection between the cuff and controller is not at an angle. 3) If a cuff does not reach or maintain the target pressure value during initial inflation, discontinue using the cuff. 4) Distribute the recall notice internally to all users. 5) If any of the affected devices have been forwarded to additional facilities, please contact these facilities and communicate this recall to them. Please list the additional facilities within the Business Reply Form 6) Following these steps, you may continue to use the product as-is. 7) Alternatively, you may return the impacted product. It is unclear when replacements will be available. 8) Complete and return the Recall Business Reply Form. It can be returned via email to SSSPFA@stryker.com Adverse reactions or quality problems experienced with the use of this product can be reported to the Complaint Hotline at 1-888-763-8803 or to http://www.stryker.com/productexperience Additional affected product was distributed, so additional recall notices were distributed to these associated customers beginning on 9/26/23.
Quantity
127,894
Distribution
US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT
Lot, serial or model codes
REF/UI-DI: 5921-030-135/885825016142, 5921-030-235/885825016166, 5921-034-135/885825016173, 5921-034-235/885825016197, 5921-218-135/885825016500, 5921-218-235/885825016296 Lots: 0000202775, 0000209215, 0000202137, 0000201380, 0000200708, 13872322, 13453772, 0000198235, 0000202066, 0000209420, 0000206425, 0000192978, 0000202577, 0000192465, 13463347, 0000204578, 0000193181, 0000198446, 0000208949, 0000195337, 0000206263, 0000193179, 13486613, 0000202777, 0000195071, 0000204159, 0000195912, 0000200905, 0000206442, 0000197869, 13594486, 0000192979, 0000200706, 0000195068, 0000201379, 0000199502, 0000199434, 0000188863, 13493683, 0000193177, 0000202065, 0000200712, 0000200242, 0000202138, 13582864, 0000195075, 13570832, 0000199808, 0000202075, 0000195340, 0000200254, 0000202932, 13774680, 13851472, 13634704, 0000199819, 0000204319, 0000204160, 0000201378, 0000200227, 0000203311, 0000206583, 13235119, 0000194243, 0000196364, 0000193841, 0000200270, 13828088, 0000196363, 13392486, 13280380, 0000192981, 0000198461, 0000196361, 0000200244, 0000203310, 0000200229, 13398036, 13352572, 0000194232, 0000200253, 0000198245, 0000202776, 0000197373, 0000187746, 13765829, 13449838, 0000200256, 0000202145, 0000198246, 0000198236, 0000204574, 0000202143, 0000194242, 13583025, 0000202074, 0000203315, 0000195073, 0000193176, 0000212764, 0000196379, 13811283, 13605685, 0000187310, 0000208444, 0000204366, 0000197867, 13865381, 13820702, 0000176566, 13663274, 0000196365, 0000210787, 0000206440, 0000193391, 0000201391, 0000193412, 0000177131, 13883084, 0000198460, 0000208345, 0000206443, 0000192466, 0000210962, 0000200230, 0000204570, 13339407, 0000204314, 0000211530, 0000209200, 0000196359, 0000210005, 0000201840, 0000205922, 0000192980, 0000205238, 0000195335, 0000206518, 0000195067, 0000203313, 0000193393, 0000206254, 0000208440, 0000202686, 0000198244, 0000203314, 0000195069, 0000208407, 0000206432, 13632837, 0000210465, 0000206505, 0000172873, 0000195338, 0000199805, 0000208321, 0000195343, 0000205241, 0000200707, 0000199807, 0000195076, 0000197375, 0000203312, 13870728, 0000202139, 0000206511, 0000193392, 13824462, 0000195070, 0000199816, 0000208410, 13871893, 0000199503, 0000206586, 13272167, 13399919, 0000195072, 0000199817, 0000192977, 0000197872, 0000186596, 0000208322, 13573489, 0000209992, 0000202684, 0000211154, 0000201837, 0000210310, 0000197870, 0000208606, 0000204571, 13883045, 0000197376, 0000194228, 0000204195, 0000198237, 0000212878, 0000209416, 0000210963, 0000210601, 0000200711, 0000194229, 0000204372, 0000196362, 13807840, 0000210008, 0000200228, 0000172314, 0000200257, 0000194230, 0000205930, 0000193839, 0000193178, 0000210313, 0000198459, 13814050, 0000202144, 0000193413, 0000195336, 13765599, 0000208951, 0000195078, 0000194227, 0000193838, 0000193840, 0000202778
510(k) numbers
not on file
PMA numbers
not on file
Product code
KCY
Firm FEI
3009350690
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register