FDA Medical Device Recalls (openFDA)
Covidien Llc: Palindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
- Recall number
- Z-1314-2022
- Recalling firm
- Covidien Llc
- Product
- Palindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
- Status
- Open, Classified
- Initiated
- 2022-06-08
- Posted
- 2022-07-07
- Terminated
- not on file
- Reason
- There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Root cause
- Process control
- Action
- Medtronic distributed a recall notification to customers by mail on 06/08/2022. Customers were instructed to immediately quarantine and discontinue use for specific lots of chronic hemodialysis catheters, which are to be returned to Medtronic. If customers have further distributed product, they are asked to forward the information from the recall notification to those recipients. All customers are asked to complete and return the Consignee Product Retrieval Confirmation Form to rs.gmbfcamitg@medtronic.com. Additionally, consignees are asked to share the notice within their organization and retain it for their records regarding the retrieval of unused product. If a consignee is aware of any adverse events associated with recalled product, they are to submit a report to FDA via Medwatch. Customers with any questions can contact their Medtronic Representative or call Customer Service at 800-962-9888, option 2. ***Updated 8/16/22*** Medtronic has created a website to help customers identify affected product: https://www.medtronic.com/covidien/en-us/products/c/renal-care-recall-lookup-tool.html#lotLookupTool
- Quantity
- 2,462 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kosovo, Kuwait, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Malta, Mayotte, Mexico, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Viet Nam, Virgin Islands, and Zimbabwe.
- Lot, serial or model codes
- Code Information organized as "Model No., Serial No., GTIN": 8888119368, 1925200090, 20884521012841; 8888119368, 1933700094, 20884521012841; 8888119368, 2001400077, 20884521012841; 8888119368, 2026000056, 20884521012841; 8888123408, 2009400114, 20884521012957; 8888128458, 1935100068, 20884521013039; 8888133508, 2026000060, 20884521013107; 8888145062, 2031800153, 10884521013407; 8888145062, 1933700095, 20884521013404; 8888145062, 2009400116, 20884521013404; 8888145062, 2031800131, 20884521013404; 8888145062, 2031800153, 20884521013404; 8888145063, 2000700056, 20884521013428; 8888145063, 2009400113, 20884521013428; 8888145063, 2027200181, 20884521013428; 8888145064, 2030100077, 20884521013442; 8888145065, 2026000067, 20884521013466; 8888145066, 2011900104, 10884521013476; 8888145066, 2019500162, 10884521013476; 8888145066, 2011900104, 20884521013473; 8888145066, 2019500162, 20884521013473
- 510(k) numbers
- ["K123196"]
- PMA numbers
- not on file
- Product code
- MSD
- Firm FEI
- 1282497
- Firm city
- Mansfield
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as