Phadia US Inc: ImmunoCAP Allergen c6, Amoxicilloyl ( Article number 14-4451-016). ImmunoCAP Specific IgE is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum or plasma; to be used with the instrument Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000.

Recall number
Z-1313-2018
Recalling firm
Phadia US Inc
Product
ImmunoCAP Allergen c6, Amoxicilloyl ( Article number 14-4451-016). ImmunoCAP Specific IgE is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum or plasma; to be used with the instrument Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000.
Status
Terminated
Initiated
2017-12-18
Posted
not on file
Terminated
2019-03-27
Reason
The stability specifications of ImmunoCAP Allergen c6, Amoxicilloyl were not met.
Root cause
Device Design
Action
Customers are being asked to please stop using all current product, and scrap any unused products. Please disregard any reported test result below the assay Limit of Quantification (LoQ 0.1 kUA/l) from the lots, as these results may not be valid. Please note that assay results above the Limit of Quantification, will only be affected by giving a lower result than the real value. Please fill in the Recall Letter Acknowledgement Form provided, including the table in the response form to obtain credit, and return the signed response by e-mail or fax.
Quantity
950 total
Distribution
Nationwide Distribution
Lot, serial or model codes
Lots  991DA, exp.2018.04.30 991DC, exp. 2018.07.31 991D7 991D9, exp.2018.02.28
510(k) numbers
["K051218","K993388"]
PMA numbers
not on file
Product code
DHB
Firm FEI
3004973408
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register