Skip to the content

FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc., dba Datascope Patient Monitoring: Datascope AS3000 Anesthesia System; Mindray, North America. The AS3000 Anesthesia Delivery System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor, provides for automatic and manual modes of ventilation, and is equipped with a monitoring system for ventilation, inspired and expired gas. The AS3000 is intended for use in operating rooms. It is used with O2, N2O and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine.

Recall number
Z-1313-2010
Recalling firm
Mindray DS USA, Inc., dba Datascope Patient Monitoring
Product
Datascope AS3000 Anesthesia System; Mindray, North America. The AS3000 Anesthesia Delivery System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor, provides for automatic and manual modes of ventilation, and is equipped with a monitoring system for ventilation, inspired and expired gas. The AS3000 is intended for use in operating rooms. It is used with O2, N2O and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine.
Status
Terminated
Initiated
2010-03-05
Posted
2010-04-08
Terminated
2012-04-25
Reason
An issue affecting the caster mount on the AS3000 System have been identified, specifically reports of the caster mount breaking, causing the wheel to fall off and the unit to tip.
Root cause
Other
Action
An "URGENT PRODUCT FIELD CORRECTIVE ACTION" letter dated March 5, 2010 was sent certified mail to all facilities/customers that have the AS3000 units. The letter described the product, problem and corrective Action/remediation by the firm. A Mindray Service Representative will contact the customers to arrange for the caster mounts to be replaced on the AS3000 units in their facilities at no cost. If you have any questions, please contact Karen Maine of the Mindray DS USA, Inc. Marketing Department at 201-995-8612 or k.maine@mindray.com.
Quantity
372 units in US, 107 internationally.
Distribution
Worldwide Distribution: USA and Mexico, Columbia, Turks & Caicos Islands, BWI; Ecuador, Bolivia, Peru, Dominican Republic, Netherlands Antilles, Chile.SA
Lot, serial or model codes
Product numbers 0998-00-3024-01.
510(k) numbers
["K080175"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register