FDA Medical Device Recalls (openFDA)

LumiraDx: Coronavirus antigen detection test system - Product Usage: intended for the qualitative detection of the nucleocapsid protein antigen to SARS-CoV-2 directly from nasal swab samples collected from individuals suspected of COVID-19 by their healthcare provider within the first twelve days of symptom onset.

Recall number
Z-1312-2021
Recalling firm
LumiraDx
Product
Coronavirus antigen detection test system - Product Usage: intended for the qualitative detection of the nucleocapsid protein antigen to SARS-CoV-2 directly from nasal swab samples collected from individuals suspected of COVID-19 by their healthcare provider within the first twelve days of symptom onset.
Status
Terminated
Initiated
2021-02-03
Posted
not on file
Terminated
2022-07-13
Reason
Two lots of test strips failed QC testing using blank buffer due to false positives.
Root cause
Process design
Action
On 2/23/2020, LumiraDx issued an Urgent Device Medical Corrective Field Action notice to customers, advising them of a mandatory field corrective action in the form of a software update for the LumiraDx SARS-CoV-2 Antigen (Ag) Test.
Quantity
2186 devices
Distribution
US nationwide distribution.
Lot, serial or model codes
catalog #: L001000330001
510(k) numbers
not on file
PMA numbers
not on file
Product code
QKP
Firm FEI
3012642695
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register