Aesculap Inc: Numerous models of nonsterile hemostatic forceps: (1) REF FB458R, Glover ATR.CLAMP CVD 215MM; (2) REF FB459R, Glover ATR.CLAMP CVD 245MM; (3) REF FB461R, Glover ATR.CLAMP 195MM; (4) REF FB462R, Glover ATR.CLAMP 210MM; (5) REF FB469R, Leland-Jones Peripheral CLMPSTR 195MM; (6) REF FB567R, Cooley Pediatric Clamp 30DG30/165MM; (7) REF FB568R, Cooley Pediatric Clamp 60DG 30/160MM; (8) REF FB569R, Cooley Pediatric Clamp 90DG30/150MM; (9) REF FB729R, Cooley Pediatric Clamp 17MM 170MM; (10) REF FB730R, Cooley Pediatric Clamp 20MM175MM; (11) REF FB736R, Cooley Pediatric Clamp 74MM 185MM; (12) REF FB737R, Cooley Pediatric Clamp 86MM 195MM; and (13) REF FB738R, Cooley Pediatric Clamp 108MM 260MM.

Recall number
Z-1311-2025
Recalling firm
Aesculap Inc
Product
Numerous models of nonsterile hemostatic forceps: (1) REF FB458R, Glover ATR.CLAMP CVD 215MM; (2) REF FB459R, Glover ATR.CLAMP CVD 245MM; (3) REF FB461R, Glover ATR.CLAMP 195MM; (4) REF FB462R, Glover ATR.CLAMP 210MM; (5) REF FB469R, Leland-Jones Peripheral CLMPSTR 195MM; (6) REF FB567R, Cooley Pediatric Clamp 30DG30/165MM; (7) REF FB568R, Cooley Pediatric Clamp 60DG 30/160MM; (8) REF FB569R, Cooley Pediatric Clamp 90DG30/150MM; (9) REF FB729R, Cooley Pediatric Clamp 17MM 170MM; (10) REF FB730R, Cooley Pediatric Clamp 20MM175MM; (11) REF FB736R, Cooley Pediatric Clamp 74MM 185MM; (12) REF FB737R, Cooley Pediatric Clamp 86MM 195MM; and (13) REF FB738R, Cooley Pediatric Clamp 108MM 260MM.
Status
Open, Classified
Initiated
2025-01-31
Posted
2025-03-11
Terminated
not on file
Reason
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
Root cause
Use error
Action
The recalling firm issued letters dated 1/29/2025 on 1/31/2025 via certified mail return-receipt. The letter explained the reason for the correction, identified the product impacted, provided the risk to health, and the actions required for the customer/user. The actions included: (1) Review the Urgent Medical Device Correction Notification in its entirety and ensure all users within the consignee organization are informed about the correction. Post the notification where the affected products are stored. If the consignee is a distributor, they are to forward the notification to their customers. The correction is to be extended to the end user/customer level; (2) Determine the current inventory of the affected items and comply with the following instructions: (a) Use the products identified for clamping of tissue and small vessels only; (b) Inspect the products for end-of-life indicators prior to use; (c) Visually inspect the products which should be clean and free of any signs of corrosion and cracks. Check the joint and jaw in the open position from both sides (if needed, use magnification; (d) Test the function of the device, including the jaw area, for deformation and loss of tension. An Acknowledgment Form was enclosed for completion and return via FAX or email within 2 weeks of receipt, even if the total inventory in the consignee location is 0.
Quantity
1,147 forceps
Distribution
Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.
Lot, serial or model codes
GTIN numbers: (1) REF FB458R - 04038653156078; (2) REF FB459R - 04038653156085; (3) REF FB461R - 04038653156108; (4) REF FB462R - 04038653156115; (5) REF FB469R - 04038653156184; (6) REF FB567R - 04038653156948; (7) REF FB568R - 04038653156955; (8) REF FB569R - 04038653156962; (9) REF FB729R - 04038653157617; (10) REF FB730R - 04038653157624; (11) REF FB736R - 04038653157686; (12) REF FB737R - 04038653157693; and (13) REF FB738R - 04038653157709.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXD
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register