Skip to the content

FDA Medical Device Recalls (openFDA)

Tekia Inc: TEKLENS II (TEKIA label) and Polylens A60 (Polytech label) Model 614, intraocular lens (IOL)

Recall number
Z-1311-06
Recalling firm
Tekia Inc
Product
TEKLENS II (TEKIA label) and Polylens A60 (Polytech label) Model 614, intraocular lens (IOL)
Status
Terminated
Initiated
2004-07-02
Posted
2006-08-02
Terminated
2012-02-10
Reason
Intraocular lenses (IOLs) with power of 17.0 D and 22.5 D are mislabeled with each other's labels.
Root cause
Other
Action
On 7/2/04, TEKIA informs first customer to stop shipment of 22.5 D IOLs from batch E0310 due to suspicion of mislabeling. On 7/8/2004 TEKIA informs first customer to stop shipping IOLs belonging to two shop orders (22.5 D E03100099-120- twenty-two (22) units, 17.0 D E100064-89- twenty-six (26) units), and requests them to retrieve products with those serial numbers from their customers. TEKIA informs second customer of suspicion of mislabeling, requests stop shipment and retrieval from their customers of the IOLs belonging to two shop orders: twelve (12) IOLs 22.5 D E03100146-153 -- eight (8) units, 22.5 D E03100218-221 --four (4) units. Return Goods Authorizations are issued.
Quantity
60 units/IOLs total
Distribution
Germany.
Lot, serial or model codes
1) Products labeled with TEKIA as manufacturer: IOLs with power of 17.0 D (4 units): 616170E031002180207, 616170E031002190207, 616170E031002200207, 616170E031002210207; IOLs with power of 22.5 D (8 units): 616225E031001460207, 616225E031001470207, 616225E031001480207, 616225E031001490207, 616225E031001500207, 616225E031001510207; 616225E031001520207, 616225E031001530207.  2) Products labeled with Polytech as manufacturer: IOLs with power of 17.0 D (26 units): 613170E031000640207, 613170E031000650207, 613170E031000660207, 613170E031000670207, 613170E031000680207, 613170E031000690207, 613170E031000700207, 613170E031000710207,613170E031000720207, 613170E031000730207, 613170E031000740207, 613170E031000750207, 613170E031000760207, 613170E031000770207, 613170E031000780207, 613170E031000790207, 613170E031000800207, 613170E031000810207, 613170E031000820207, 613170E031000830207, 613170E031000840207, 613170E031000851207, 613170E031000860207, 613170E031000870207, 613170E031000880207, 613170E031000890207; IOLs with power of 22.5 D (22 units): 613225E031000990207, 613225E031001000207, 613225E031001010207, 613225E031001020207, 613225E031001030207, 613225E031001040207, 613225E031001050207, 613225E031001060207, 613225E031001070207, 613225E031001080207, 613225E031001090207, 613225E031001100207, 613225E031001110207, 613225E031001120207, 613225E031001130207, 613225E031001140207, 613225E031001150207, 613225E031001160207, 613225E031001170207, 613225E031001180207, 613225E031001190207, 613225E031001200207.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HQL
Firm FEI
3002023310
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register