FDA Medical Device Recalls (openFDA)
Tekia Inc: TEKLENS II (TEKIA label) and Polylens A60 (Polytech label) Model 614, intraocular lens (IOL)
- Recall number
- Z-1311-06
- Recalling firm
- Tekia Inc
- Product
- TEKLENS II (TEKIA label) and Polylens A60 (Polytech label) Model 614, intraocular lens (IOL)
- Status
- Terminated
- Initiated
- 2004-07-02
- Posted
- 2006-08-02
- Terminated
- 2012-02-10
- Reason
- Intraocular lenses (IOLs) with power of 17.0 D and 22.5 D are mislabeled with each other's labels.
- Root cause
- Other
- Action
- On 7/2/04, TEKIA informs first customer to stop shipment of 22.5 D IOLs from batch E0310 due to suspicion of mislabeling. On 7/8/2004 TEKIA informs first customer to stop shipping IOLs belonging to two shop orders (22.5 D E03100099-120- twenty-two (22) units, 17.0 D E100064-89- twenty-six (26) units), and requests them to retrieve products with those serial numbers from their customers. TEKIA informs second customer of suspicion of mislabeling, requests stop shipment and retrieval from their customers of the IOLs belonging to two shop orders: twelve (12) IOLs 22.5 D E03100146-153 -- eight (8) units, 22.5 D E03100218-221 --four (4) units. Return Goods Authorizations are issued.
- Quantity
- 60 units/IOLs total
- Distribution
- Germany.
- Lot, serial or model codes
- 1) Products labeled with TEKIA as manufacturer: IOLs with power of 17.0 D (4 units): 616170E031002180207, 616170E031002190207, 616170E031002200207, 616170E031002210207; IOLs with power of 22.5 D (8 units): 616225E031001460207, 616225E031001470207, 616225E031001480207, 616225E031001490207, 616225E031001500207, 616225E031001510207; 616225E031001520207, 616225E031001530207. 2) Products labeled with Polytech as manufacturer: IOLs with power of 17.0 D (26 units): 613170E031000640207, 613170E031000650207, 613170E031000660207, 613170E031000670207, 613170E031000680207, 613170E031000690207, 613170E031000700207, 613170E031000710207,613170E031000720207, 613170E031000730207, 613170E031000740207, 613170E031000750207, 613170E031000760207, 613170E031000770207, 613170E031000780207, 613170E031000790207, 613170E031000800207, 613170E031000810207, 613170E031000820207, 613170E031000830207, 613170E031000840207, 613170E031000851207, 613170E031000860207, 613170E031000870207, 613170E031000880207, 613170E031000890207; IOLs with power of 22.5 D (22 units): 613225E031000990207, 613225E031001000207, 613225E031001010207, 613225E031001020207, 613225E031001030207, 613225E031001040207, 613225E031001050207, 613225E031001060207, 613225E031001070207, 613225E031001080207, 613225E031001090207, 613225E031001100207, 613225E031001110207, 613225E031001120207, 613225E031001130207, 613225E031001140207, 613225E031001150207, 613225E031001160207, 613225E031001170207, 613225E031001180207, 613225E031001190207, 613225E031001200207.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HQL
- Firm FEI
- 3002023310
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28