Elekta Instrument AB: Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund Catheter Insertion Needle Kit III; (2) Part #A2800-15, Backlund Catheter Insertion Needle Kit I; (3) Part #A2600-01, Backlund Haematoma Evacuator Kit; (4) Part #A2200-01, Salcman Twist Drill Kit I; (5) Part #907801, Insertion Cannula Kit; (6) Part #50398-01, Catheter Inserter Needle 1,5; (7) Part #307165, Insertion Cannula 190 MM; (8) Part #60377-02, Twist Drill 3.2 MM; (9) Part #60377-01, Twist Drill 2.1 MM; (10) Part #50376-01, Salcman Twist Drill 2.1/3.2; (11) Part #14001050, Catheter Inserter Needle 2,6; and (12) Part #1002248, Frame G w Straight/Curved Front.
- Recall number
- Z-1309-2023
- Recalling firm
- Elekta Instrument AB
- Product
- Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund Catheter Insertion Needle Kit III; (2) Part #A2800-15, Backlund Catheter Insertion Needle Kit I; (3) Part #A2600-01, Backlund Haematoma Evacuator Kit; (4) Part #A2200-01, Salcman Twist Drill Kit I; (5) Part #907801, Insertion Cannula Kit; (6) Part #50398-01, Catheter Inserter Needle 1,5; (7) Part #307165, Insertion Cannula 190 MM; (8) Part #60377-02, Twist Drill 3.2 MM; (9) Part #60377-01, Twist Drill 2.1 MM; (10) Part #50376-01, Salcman Twist Drill 2.1/3.2; (11) Part #14001050, Catheter Inserter Needle 2,6; and (12) Part #1002248, Frame G w Straight/Curved Front.
- Status
- Open, Classified
- Initiated
- 2023-03-01
- Posted
- 2023-03-30
- Terminated
- not on file
- Reason
- The incorrect IFU was distributed with the devices.
- Root cause
- Labeling Change Control
- Action
- The firm issued a Field Safety Notification dated February 2023 on 3/1/2023 via email. The notification describes the problem, lists the part numbers involved, the clinical impact, and recommended user actions also letting them know the correct IFU will be shipped to all affected customers. The recommended user actions include posting the notice in a place accessible to all users until this action is closed and advising the appropriate personnel working with the product on the content of the letter. The consignee is informed that Elekta will provide customers with the correct IFU. An Acknowledgement Form is enclosed to acknowledge receipt of the notification to Elekta Care Community or complete the form and return it to Elekta no later than within 30 days. The consignee is to provide the device serial number and location or site and sign the form that they had read and understand the notice. On 3/2/2023, the correct IFU version was released to all internal service managers who will ensure all customers received the corrected version.
- Quantity
- 70 devices
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, FL, GA, ID, IL, KS, MA, MD, MI, MN, MO, MS, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. There was also government distribution.
- Lot, serial or model codes
- UDI-DI: (1) 07340048301305; (2) 07340048300285; (3) 07340048301220; (4) 07340048301145; (5) 07340048300315; (6) 07340048308618; (7) 07340048308632; (8) 07340048308571; (9) 07340048308564; (10 07340048308601; (11) 07340048308625; and (12) 07340048306348.
- 510(k) numbers
- K152558
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 3002806883
- Firm city
- Stockholm
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28