Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta Instrument AB: Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund Catheter Insertion Needle Kit III; (2) Part #A2800-15, Backlund Catheter Insertion Needle Kit I; (3) Part #A2600-01, Backlund Haematoma Evacuator Kit; (4) Part #A2200-01, Salcman Twist Drill Kit I; (5) Part #907801, Insertion Cannula Kit; (6) Part #50398-01, Catheter Inserter Needle 1,5; (7) Part #307165, Insertion Cannula 190 MM; (8) Part #60377-02, Twist Drill 3.2 MM; (9) Part #60377-01, Twist Drill 2.1 MM; (10) Part #50376-01, Salcman Twist Drill 2.1/3.2; (11) Part #14001050, Catheter Inserter Needle 2,6; and (12) Part #1002248, Frame G w Straight/Curved Front.

Recall number
Z-1309-2023
Recalling firm
Elekta Instrument AB
Product
Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund Catheter Insertion Needle Kit III; (2) Part #A2800-15, Backlund Catheter Insertion Needle Kit I; (3) Part #A2600-01, Backlund Haematoma Evacuator Kit; (4) Part #A2200-01, Salcman Twist Drill Kit I; (5) Part #907801, Insertion Cannula Kit; (6) Part #50398-01, Catheter Inserter Needle 1,5; (7) Part #307165, Insertion Cannula 190 MM; (8) Part #60377-02, Twist Drill 3.2 MM; (9) Part #60377-01, Twist Drill 2.1 MM; (10) Part #50376-01, Salcman Twist Drill 2.1/3.2; (11) Part #14001050, Catheter Inserter Needle 2,6; and (12) Part #1002248, Frame G w Straight/Curved Front.
Status
Open, Classified
Initiated
2023-03-01
Posted
2023-03-30
Terminated
not on file
Reason
The incorrect IFU was distributed with the devices.
Root cause
Labeling Change Control
Action
The firm issued a Field Safety Notification dated February 2023 on 3/1/2023 via email. The notification describes the problem, lists the part numbers involved, the clinical impact, and recommended user actions also letting them know the correct IFU will be shipped to all affected customers. The recommended user actions include posting the notice in a place accessible to all users until this action is closed and advising the appropriate personnel working with the product on the content of the letter. The consignee is informed that Elekta will provide customers with the correct IFU. An Acknowledgement Form is enclosed to acknowledge receipt of the notification to Elekta Care Community or complete the form and return it to Elekta no later than within 30 days. The consignee is to provide the device serial number and location or site and sign the form that they had read and understand the notice. On 3/2/2023, the correct IFU version was released to all internal service managers who will ensure all customers received the corrected version.
Quantity
70 devices
Distribution
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, ID, IL, KS, MA, MD, MI, MN, MO, MS, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. There was also government distribution.
Lot, serial or model codes
UDI-DI: (1) 07340048301305; (2) 07340048300285; (3) 07340048301220; (4) 07340048301145; (5) 07340048300315; (6) 07340048308618; (7) 07340048308632; (8) 07340048308571; (9) 07340048308564; (10 07340048308601; (11) 07340048308625; and (12) 07340048306348.
510(k) numbers
K152558
PMA numbers
not on file
Product code
HAW
Firm FEI
3002806883
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register