Skip to the content

FDA Medical Device Recalls (openFDA)

National Biological Corp: HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for diagnosed skin disorders.

Recall number
Z-1309-2008
Recalling firm
National Biological Corp
Product
HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for diagnosed skin disorders.
Status
Terminated
Initiated
2008-02-19
Posted
2008-08-28
Terminated
2011-09-20
Reason
Software allows operator to override "low line voltage" error warning and store light intensity value.
Root cause
Software design
Action
National Biological sent an "Important safety reminders for your Houva DermaSense phototherapy device" letters, dated 2/11/08 to users. It states that the user should collect a "Profile" of the light energy and record it on a daily/weekly flow chart. The chart should be in plain view so each operator can clearly determine the units output. If an error message appears in the display panel, treatment should not be continued unless the phototherapy booth operator was able to identify and correct the cause of the error. Make sure both breakers are in the "ON" position before starting up the unit each morning and before giving treatment.
Quantity
48 Units
Distribution
Worldwide distribution --- USA including states of: CA, DE, FL, GA, ID, IL, IN, MA, MI, MN, NC, NH, NY, OH, OR, PA, TX, VA, WA and WI, and country of Canada.
Lot, serial or model codes
Serial #'s: 10134, 10145, 10147, 10148, 10105, 10151, 10112, 10113, 10114, 10101, 10150, 10140, 10103, 10123, 10135, 10135, 10138, 10137, 10122
510(k) numbers
["K041212"]
PMA numbers
not on file
Product code
FTC
Firm FEI
1521608
Firm city
Beachwood
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register