FDA Medical Device Recalls (openFDA)

AGFA Corp.: Impax 4.5, Impax 5.0, Impax 5.2 and Impax 5.2x Picture Archiving Systems (PACs) with integrated Voxar 3D software; and Impax 5.0 and Impax 5.2 PACs with integrated MediCAD software. The softwares are used to provide digital X-ray images for pre-surgical orthopedic planning. AGFA Corp., Greenville, South Carolina 29601

Recall number
Z-1307-2008
Recalling firm
AGFA Corp.
Product
Impax 4.5, Impax 5.0, Impax 5.2 and Impax 5.2x Picture Archiving Systems (PACs) with integrated Voxar 3D software; and Impax 5.0 and Impax 5.2 PACs with integrated MediCAD software. The softwares are used to provide digital X-ray images for pre-surgical orthopedic planning. AGFA Corp., Greenville, South Carolina 29601
Status
Terminated
Initiated
2007-02-15
Posted
2008-07-02
Terminated
2009-06-30
Reason
Misidentification: The orthopedic planning X-ray images for one patient are misidentified as the images for another patient.
Root cause
Software design
Action
On 2/15/07, AGFA notified their consignees of the situation by sending them "Urgent Medical Device Correction" letters, instructing them on how to work around the problem until the firm was able to install the upgraded software that corrected the problem. On 12/11/2007, the firm reported that the upgraded software had been installed at all affected sites.
Quantity
879 units
Distribution
Worldwide Distribution - USA including states of AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, KS, MA, MD, ME, MI, NC, NJ, NV, NY, NV, OH, OK, PA, SC, TN, TX, UT, VA and WA, and country of Canada.
Lot, serial or model codes
Hectec MediCAD: ETSAN; VOXAR 3D: L72JO; and Impax 4.5 or 5.0 or 5.2 Client /OT3000: A component of Order Codes: 60+00021533, 60+00021534, 60+00022471, 60+00022472, 60+00015451, 60+00015453, 60+00021535 and 60+00024519.
510(k) numbers
["K022292"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register