Abbott Diabetes Care, Inc.: Abbott brand FreeStyle Flash Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics International, Long Jing Industrial Estate, Long Jing Road, Xi Li Town, Nantou District, Shenzhen, China
- Recall number
- Z-1307-06
- Recalling firm
- Abbott Diabetes Care, Inc.
- Product
- Abbott brand FreeStyle Flash Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics International, Long Jing Industrial Estate, Long Jing Road, Xi Li Town, Nantou District, Shenzhen, China
- Status
- Terminated
- Initiated
- 2006-05-22
- Posted
- 2006-08-03
- Terminated
- 2006-12-21
- Reason
- The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration data including selectable unit of measure, and strip calibration code.
- Root cause
- Other
- Action
- On 5/22/06, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions for the corrective actions.
- Quantity
- 2,365,000 meters (approx.-both types
- Distribution
- Worldwide
- Lot, serial or model codes
- All codes.
- 510(k) numbers
- ["K051839","K992684"]
- PMA numbers
- not on file
- Product code
- NBW
- Firm FEI
- 3002803444
- Firm city
- Alameda
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28