Skip to the content

FDA Medical Device Recalls (openFDA)

Fujirebio Diagnostics, Inc.: Product Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma Calibrators: The -Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G -Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 1.5 mL (4 Concentrations) CAL 1 0 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL -Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Recall number
Z-1306-2026
Recalling firm
Fujirebio Diagnostics, Inc.
Product
Product Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma Calibrators: The -Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G -Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 1.5 mL (4 Concentrations) CAL 1 0 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL -Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Status
Open, Classified
Initiated
2025-12-11
Posted
2026-02-05
Terminated
not on file
Reason
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Root cause
Under Investigation by firm
Action
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.
Quantity
128 units
Distribution
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Lot, serial or model codes
Model/Catalog Number: 235454; UDI-DI: 04987270235454; Lot Numbers: D8C6022U, D8C6044U;
510(k) numbers
["K242706"]
PMA numbers
not on file
Product code
SET
Firm FEI
2521625
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register