FDA Medical Device Recalls (openFDA)
Penner Mfg Inc: Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.
- Recall number
- Z-1306-2015
- Recalling firm
- Penner Mfg Inc
- Product
- Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.
- Status
- Terminated
- Initiated
- 2015-01-30
- Posted
- 2015-03-20
- Terminated
- 2016-03-10
- Reason
- If the patient sits on either wing, not squarely on the center section of the Stretcher Lift, tipping and injury to the operator or patient could result. The device manual was updated with an additional warning statement.
- Root cause
- Use error
- Action
- Beginning on 1/30/2015, the recalling firm sent letters via certified mail to their customers informing them of the change to the stretcher manual along with a revised manual which incorporated those changes. For questions regarding this recall call 402-694-5003.
- Quantity
- 158
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada.
- Lot, serial or model codes
- Serial numbers: 01050741201, 01071198001, 01071198002, 01071198003, 01071198004, 01091136401, 01091136402, 01091136403, 01101314201, 01101322801, 01111505901, 02071228401, 02071228402, 02071228403, 02071228404, 02080975001, 02080975002, 02091137801, 02091149801, 02121713401, 03050791101, 03050804301, 03050804302, 03050804303, 03050804304, 03050804305, 03050804306, 03071247201, 03071247202, 03071247203, 03071247204, 03080984601, 03080984602, 03080984603, 03091161601, 03101347201, 03121742801, 03122862901, 04040581301, 04040581302, 04040581303, 04040581304, 04040581305, 04040581306, 04040581307, 04040581308, 04040581310, 04040581311, 04040581312, 04040581313, 04040581314, 04040586101, 04050810401, 04050810402, 04050810403, 04050810404, 04050810406, 04050810408, 04050810410, 04061028101, 04061028102, 04061028103, 04061028104, 04061028105, 04081009901, 04091187201, 04111563801, 04111568501, 05071272201, 05071272202, 05071272203, 05071272204, 05071272205, 05081019501, 05091197901, 05091197902, 05091201301, 05111513201, 05111576501, 05121762001, 06050840801, 06050840802, 06050840803, 06050840804, 06050840805, 06050840806, 06081037501, 06081039402, 06081039403, 06101394201, 07071320201, 07071320202, 07071328501, 07071328502, 07071328503, 07091223901, 07091223902, 07101410901, 07101415901, 08040660501, 08040660502, 08061090801, 08061090802, 08061090803, 08061090804, 08101430502, 08101439601, 09081073301, 09081074401, 09081076501, 09101453401, 09101457101, 09111641101, 09111641301, 10061121201, 10061121202, 10061121203, 10061121204, 10071396601, 10071396602, 10071396603, 10071396604, 10101463201, 10101464701, 10111652001, 11040710501, 11040710502, 11040710503, 11040710504, 11040710505, 11040710506, 11040710507, 11040710508, 11040710601, 11040710602, 11040710603, 11040710604, 11040710605, 11040710606, 11040710607, 11040710608, 11081101401, 11081105901, 11081113501, 11091285601, 11091285602, 11101475201, 11101476201, 11101476401, 11101477901, 11101478101, 12101495201, 12111677201, 12111686301, 12111686302, 12111688401, 03121734201S, 04040581301A
- 510(k) numbers
- ["K962288"]
- PMA numbers
- not on file
- Product code
- MMP
- Firm FEI
- 1922538
- Firm city
- Aurora
- Firm state
- NE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28