Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Point Of Care Inc.: i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.

Recall number
Z-1306-2014
Recalling firm
Abbott Point Of Care Inc.
Product
i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Status
Terminated
Initiated
2014-01-01
Posted
2014-03-27
Terminated
2014-10-20
Reason
Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
Root cause
Process control
Action
APOC sent an Urgent Recall Notice dated January 2014 via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product and provide a copy of this letter to any institution that was forwarded the affected product. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card. To ensure continuity of supply at their facility, customers were instructed to follow their standard process to place an order or call Customer Sedrvice at 1-800-323-9100, Option #2, then Option #9). Customers with questions were instructed to contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1 of their Abbott Point of Care representative. For questions regarding this recall call 800-323-9100.
Quantity
48, 450 US, 1400 OUS (49,850 catridges)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Lot, serial or model codes
List number 03P78-25 :Lot number N13183
510(k) numbers
["K940918"]
PMA numbers
not on file
Product code
CHL
Firm FEI
2245578
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register