Skip to the content

FDA Medical Device Recalls (openFDA)

Natus Neurology Inc: TENSIVE Conductive Adhesive Gel, REF 016-401600

Recall number
Z-1305-2024
Recalling firm
Natus Neurology Inc
Product
TENSIVE Conductive Adhesive Gel, REF 016-401600
Status
Open, Classified
Initiated
2024-02-07
Posted
2024-03-19
Terminated
not on file
Reason
Natus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five complaints. These customers reported that the gel is too thick, is discolored (darker than usual), and in one instance, patients reported irritation after procedures using the gel.
Root cause
Under Investigation by firm
Action
Natus issued an Urgent Medical Device Recall Notice to its consignees on 02/07/2024 via UPS and email. The notice explained the issue and risk to health. It also requested the consignee cease use, complete and return the form for the replacement of the product.
Quantity
80 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
UDI/DI 00855683006333, Lot Numbers: A0623009
510(k) numbers
["K852956"]
PMA numbers
not on file
Product code
GYB
Firm FEI
3010611950
Firm city
Middleton
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register