Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software

Recall number
Z-1304-2025
Recalling firm
CareFusion 303, Inc.
Product
BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software
Status
Open, Classified
Initiated
2025-02-18
Posted
2025-03-13
Terminated
not on file
Reason
Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
Root cause
Under Investigation by firm
Action
On February 18, 2025, Becton, Dickinson issued a " Urgent Medical Device Product Correction" Notification. to affected consignees via Fed-Ex and E-Mail. 1. Users should review all infusion parameters pre-populated and confirm they are correct before pressing START as stated in the BD Alaris User Manual. 2. Disseminate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 3. BD Project Managers will engage with you to address any questions regarding this notice, and specific next steps for your facility. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form, regardless of whether you have any of the impacted material. This will allow BD to confirm receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.
Quantity
not on file
Distribution
US: AZ, CA, DC, FL, GA, IL, IN, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI OUS: None
Lot, serial or model codes
Software v12.5.1 or v12.5.2, and prior versions (4.33, 12.0.1, 12.0.2, 12.1, 12.1.2, 12.3) UDIs: 10885403960123 10885403519666 (10885403960116)
510(k) numbers
["K211218"]
PMA numbers
not on file
Product code
PHC
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register