FDA Medical Device Recalls (openFDA)
CareFusion 303, Inc.: BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software
- Recall number
- Z-1304-2025
- Recalling firm
- CareFusion 303, Inc.
- Product
- BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software
- Status
- Open, Classified
- Initiated
- 2025-02-18
- Posted
- 2025-03-13
- Terminated
- not on file
- Reason
- Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
- Root cause
- Under Investigation by firm
- Action
- On February 18, 2025, Becton, Dickinson issued a " Urgent Medical Device Product Correction" Notification. to affected consignees via Fed-Ex and E-Mail. 1. Users should review all infusion parameters pre-populated and confirm they are correct before pressing START as stated in the BD Alaris User Manual. 2. Disseminate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 3. BD Project Managers will engage with you to address any questions regarding this notice, and specific next steps for your facility. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form, regardless of whether you have any of the impacted material. This will allow BD to confirm receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.
- Quantity
- not on file
- Distribution
- US: AZ, CA, DC, FL, GA, IL, IN, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI OUS: None
- Lot, serial or model codes
- Software v12.5.1 or v12.5.2, and prior versions (4.33, 12.0.1, 12.0.2, 12.1, 12.1.2, 12.3) UDIs: 10885403960123 10885403519666 (10885403960116)
- 510(k) numbers
- ["K211218"]
- PMA numbers
- not on file
- Product code
- PHC
- Firm FEI
- 2016493
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as