Skip to the content

FDA Medical Device Recalls (openFDA)

Dynavox Systems Llc: Dynavox T10 Speech Generating Device ( including Batteries)

Recall number
Z-1304-2018
Recalling firm
Dynavox Systems Llc
Product
Dynavox T10 Speech Generating Device ( including Batteries)
Status
Terminated
Initiated
2017-12-14
Posted
not on file
Terminated
2019-12-11
Reason
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
Root cause
Component design/selection
Action
On February 26, 2018 an URGENT MEDICAL RECALL LETTER was issued to customers with the subject line listing the serial number specific to that customer along with the date of purchase. The letter informs customers of the issue and instructs customers to return product to one of their facilities to have the product repaired until an RA. Questions or concerns can be directed to 1-800-344-1778, Press 1 and then Press 2.
Quantity
4615
Distribution
Nationally
Lot, serial or model codes
TS1308 TS1411  Part Numbers: T10: 750394  Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.
510(k) numbers
not on file
PMA numbers
not on file
Product code
ILQ
Firm FEI
3004959902
Firm city
Pittsburgh
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register