Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems LLC: GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.

Recall number
Z-1304-06
Recalling firm
General Electric Med Systems LLC
Product
GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.
Status
Terminated
Initiated
2006-05-15
Posted
2006-09-22
Terminated
2007-05-25
Reason
Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in position to intercept the entire cross section of the useful beam
Root cause
Other
Action
An Important Notice for Precision 500D, dated 11/24/2005 was hand delivered to all customers beginning 05/15/06, by GE field service engineers at the time they were to perform the correction to the device.
Quantity
1,289
Distribution
Worldwide, including USA, Puerto Rico, Canada, France, Germany, Indonesia, Italy, Japan, Jordan, Kuwait, Macedonia, Panama, Russia, Saudi Arabia, Singapore, and United Kingdom.
Lot, serial or model codes
All serial numbers affected
510(k) numbers
["K011624"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register