Skip to the content

FDA Medical Device Recalls (openFDA)

Biocardia, Inc.: Morph AccessPro Steerable Introducer; 6F Introducer, 90 cm working length: Single use only. BioCardia, Inc. San Carlos, CA 94070 BioCardia Morph AccessPro is intended to provide a pathway through medical instruments cush as balloon dilation catheters, guidewires or other therapeutic devices that may be introduced into the peripheral vasculature.

Recall number
Z-1302-2014
Recalling firm
Biocardia, Inc.
Product
Morph AccessPro Steerable Introducer; 6F Introducer, 90 cm working length: Single use only. BioCardia, Inc. San Carlos, CA 94070 BioCardia Morph AccessPro is intended to provide a pathway through medical instruments cush as balloon dilation catheters, guidewires or other therapeutic devices that may be introduced into the peripheral vasculature.
Status
Terminated
Initiated
2014-03-03
Posted
2014-03-27
Terminated
2014-06-19
Reason
Firm discovered the potential for particulates in the inner lumen of Morph Access Pro Steerable Introducer, which could result in am embolic event.
Root cause
Other
Action
The firm, BioCardia, Inc., sent an "URGENT-VOLUNTARY MEDICAL DEVICE RECALL" letter dated March 3, 2014 its customers, hospital's Director of Risk Management, via USPS mail and priority mail to Cardiac Catheterization Laboratory. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your inventory; quarantine the product; return the product to BioCardia and contact Customer Service at 650-226-0148 to coordinate product return; in the event you no longer have this device as a result of further distribution, identify the purchaser and notify them at once of this recall; and complete and return the enclosed MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form via FAX to 650-631-3731 or mail to: BioCardia, Inc., 125 Shoreway Road, Suite B, San Carlos, CA 94070 as soon as possible. If you have any questions, call the Vice President, Quality Assurance at 650-226-0143.
Quantity
~2541 units - both catalog numbers.
Distribution
Worldwide Distribution: US (nationwide) and country of: Israel.
Lot, serial or model codes
Catalog number MAP690 -lot numbers:  1051  1063  1083  1106  1119  1137  1138  1140  1146  1148  1151  1154  1160  1162  1183  1186
510(k) numbers
["K090999"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3003610571
Firm city
San Carlos
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register