Skip to the content

FDA Medical Device Recalls (openFDA)

Volcano Corp: Volcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 2797

Recall number
Z-1301-2022
Recalling firm
Volcano Corp
Product
Volcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 2797
Status
Open, Classified
Initiated
2022-05-13
Posted
not on file
Terminated
not on file
Reason
Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications
Root cause
Process control
Action
On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions: 1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal. 2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy. 3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred. 4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email: igtdc.r@philips.com
Quantity
14.031 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.
Lot, serial or model codes
All Lots containing IFU Part Number 300004856772/A Revision Date: 05/2021 , ; UDI/DI: 00845225002329
510(k) numbers
["K121273"]
PMA numbers
not on file
Product code
OBJ
Firm FEI
3008363989
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register