FDA Medical Device Recalls (openFDA)
Volcano Corp: Volcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 2797
- Recall number
- Z-1301-2022
- Recalling firm
- Volcano Corp
- Product
- Volcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 2797
- Status
- Open, Classified
- Initiated
- 2022-05-13
- Posted
- not on file
- Terminated
- not on file
- Reason
- Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications
- Root cause
- Process control
- Action
- On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions: 1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal. 2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy. 3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred. 4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email: igtdc.r@philips.com
- Quantity
- 14.031 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.
- Lot, serial or model codes
- All Lots containing IFU Part Number 300004856772/A Revision Date: 05/2021 , ; UDI/DI: 00845225002329
- 510(k) numbers
- ["K121273"]
- PMA numbers
- not on file
- Product code
- OBJ
- Firm FEI
- 3008363989
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28